 13

EN

KSL-Z by KEELER

6.2  PREPARING THE PATIENT AND USING THE SLIT LAMP

Parts of the slit lamp coming in to contact with the patient should be 

cleaned in accordance with these instructions prior to the examination. 

Keeler recommends the use of disposable hygienic chinrest tissues on the 

Chinrest before the patients place their chin on it.
Never use the instrument if visibly damaged and periodically inspect it for 

signs of damage or misuse.

1.  The patient should be as comfortable as possible and with the patient in the chinrest 

adjust the chinrest height (6) so that the patient’s eyes are level with the height marking 

(3) on the chinrest support.

2.  Focus the eyepieces using the test bar as described earlier, and if you have not already 

done so set them to your interpupillary distance by holding both eyepiece bodies and 

rotating them inwards or outwards until they are correct for your PD.

3.  Switch on the illumination, making sure the rheostat (21) is set to a low level to 

minimise the patient’s exposure to light hazard.

4.  Rotate the joystick (20) until the light beam is at eye level.
5.  Holding the joystick vertical, move the slit Lamp base towards the patient until the slit 

beam appears focused on the patient’s cornea.

6.  Adjust slit width (14), magnifi cation (27), slit rotation (13) and slit angle etc. as required 

to perform the examination.

7.  To offset the slit for sclerotic scatter or retro illumination, rotate the illumination prism 

assembly (11) to the left or right as desired.

8.  When using the blue fi lter (12) the user may wish to insert the yellow barrier fi lter (9). 

The yellow barrier fi lter is out when the knob is up, in when it is down.

9.  When the examination is complete, set the rheostat to a low level and switch off the Slit 

Lamp.

20

21

9

12

27

11

13

14

3

31

Содержание Z Series

Страница 1: ...SLIT LAMP INSTRUCTIONS FOR USE Z Series A world without vision loss...

Страница 2: ......

Страница 3: ...OR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 15 7 2...

Страница 4: ...ty Humidity limitation Use by date Serial number Catalogue number Medical device Translation The Keeler Slit Lamp is designed and built in conformity with Directive 93 42 EEC Regulation EU 2017 745 an...

Страница 5: ...sembly and a table top with power supply and accessory drawer The light intensity is controlled by a variable rheostat located on the XYZ translation base There are a number of selectable filters allo...

Страница 6: ...y the specialized technicians of the manufacturer s Technical Service Centre or by personnel trained and authorised by the manufacturer The manufacturer declines any and all responsibility for loss an...

Страница 7: ...ortant when the unit has been stored or transported in a cold environment this can cause severe condensation to develop on the optical elements 2 3 CONTRAINDICATION There is no restriction to patient...

Страница 8: ...on light 5 Chinrest 6 Chinrest height adjuster 7 Patient grab handles 5 3 4 1 2 7 6 13 14 15 8 16 17 18 19 11 10 12 9 KSL Z Series 8 Eyepiece assembly securing knob 9 Yellow filter knob up out 10 Illu...

Страница 9: ...Runner covers 25 Z type tonometer mounting hole 26 Yellow filter knob up out 27 Lock for securing magnification body 28 Magnification change drum 29 Eyepieces adjustable for PD and dioptre correction...

Страница 10: ...s castors ensure the following before moving it to another location 1 The table is at its lowest position 2 The power cord is removed 3 The Slit Lamp arm and base locking knobs are tightened 4 The run...

Страница 11: ...Lamp base with the base notch b and pin c aligned Attach the tower to the base using the hex bolt removed earlier and tighten using the wrench provided 2 Carefully fit the microscope body to the arm e...

Страница 12: ...le power supply see section 9 4 Power Supply page 20 2 If the Keeler Slit Lamp is used with a power supply or cables other than those supplied this may result in increased emissions or decreased immun...

Страница 13: ...in use the Tonometer should be removed from the Slit Lamp and placed securely back in the packaging or a suitable location Applanation Tonometer Keeler Fixed Z TYPE This instrument is for those who wi...

Страница 14: ...he slit to full width 14 set the magnification to x16 27 3 Adjust the eyepieces pupillary distance by holding both eyepiece bodies and rotating them inwards or outwards until they are correct for your...

Страница 15: ...istance by holding both eyepiece bodies and rotating them inwards or outwards until they are correct for your PD 3 Switch on the illumination making sure the rheostat 21 is set to a low level to minim...

Страница 16: ...e 0 parallel eyepieces 38mm 85mm Convergent angle of optical axis 13 Magnification change 5 step drum 3 step drum Magnification Field of view 3 5 6x 34mm 3 3 10x 22mm 3 3 16x 14mm 3 3 25x 8 5mm 3 5 40...

Страница 17: ...with a soft clean lens cloth 2 Care must be taken to keep the objective and the eyepiece lenses clean use only soft clean lens cloths to clean optical surfaces 7 4 ELECTRICAL CONNECTIONS 1 Routinely c...

Страница 18: ...g or using this instrument please read carefully and observe what is described here Portable or mobile type radio frequency communication units may have an adverse effect on this instrument resulting...

Страница 19: ...to line s 2 kV line s for input output line s 1 kV line s to line s 2 kV line s for input output line s Mains power quality should be that of a typical professional healthcare facility Voltage dips sh...

Страница 20: ...objects and people 1 Field strengths from fixed transmitters such as base stations cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot b...

Страница 21: ...distance for the higher frequency applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 9 4 T...

Страница 22: ...000 6 2 EN IEC 61000 6 3 Fuse T2 5AH 250V Power supply output 12V DC 2 5 amps must be EN IEC 60601 compliant Complies with Electrical Safety Medical EN IEC 60601 1 Electromagnetic compatibility EN IEC...

Страница 23: ...e and transportation Do not store or transport this instrument in conditions where the temperature may rise above 50 C 10 ACCESSORIES AND SPARES Item Part Number Description Kapture Imaging Software L...

Страница 24: ...ates that this product shall not be treated as household waste To reduce the environmental impact of WEEE Waste Electrical Electronic Equipment and minimise the volume of WEEE entering landfills we en...

Страница 25: ......

Страница 26: ...2021 USA Sales Office Keeler USA 3222 Phoenixville Pike Building 50 Malvern PA 19355 USA Toll Free 1 800 523 5620 Tel 1 610 353 4350 Fax 1 610 353 7814 India Office Keeler India Halma India Pvt Ltd P...

Отзывы: