Technical Information
Back Next
Home
It is well established that exposure of the eye to intense light
sources for extended periods of time poses a risk of retinal photic
injury. Many ophthalmic instruments illuminate the eye with
intense light. The decision about the intensity of the light level
to use in any procedure must be made on a case to case basis. In
each case, the clinician must take a risk benefit judgement about
the intensity of light to be used. Use of insufficient intensity may
result in inadequate visualization and in adverse effects more
serious than retinal photic damage. Further, despite all efforts
taken to minimise the risk of retinal damage, damage may still
occur. Retinal photic injury is a possible complication of the need
to use bright light clearly visualize ocular structure during
delicate ophthalmic surgical procedure.
While no visible retinal lesions have been identified for
ophthalmic instruments, it is recommended that illumination
levels be set to the minimum level necessary to perform the
diagnostic function. Young children and persons with diseased
eyes may be at a higher risk. The risk may also be increased if
the person being examined has had any exposure with the same
instrument or any other ophthalmic instrument using an intense
visible light source during the previous 24 hours. This will apply
particularly if the eye has been exposed to retinal photography.
0
2
4
6
8
10
12
14
16
300
400
500
600
700
800
900 1000 1100
Wavelength / nm
Irradiance / mW m-2 nm-1
Spectral irradiance of Vantage Plus indirect ophthalmoscope at user plane.