4. Safety

Warnings and cautions – General (cont’d)

•  The product has been designed to function safely when at an  
  ambient temperature b10°C and +35°C

•  Keep out of the reach of children 

•  To prevent condensation from forming, allow instrument to  
  come to room temperature before use 

•  For indoor use only (protect from moisture)

•  There are no user serviceable parts inside. Contact authorised  
  service representative for further information

•  Ensure device is securely held in docking station to minimise  
  risk of injury or damage to equipment

•  Follow guidance on cleaning / routine maintenance to prevent  
  personal injury / damage to equipment

•  Failure to carry out recommended routine maintenance as per  
  the instructions in this IFU may reduce the operational lifetime 
  of the product

 

•  At product end of life dispose of in accordance with local    
  environmental guidelines (WEEE)

•  To isolate the equipment, disconnect from the mains or switch  
  off at the mains

Chargers

   

Do not fit mains power adapter into a damaged mains

   

outlet socket.

   

Route power cords safely to eliminate risk of tripping    

   

or damage to equipment.

Only Keeler handles with a red base can be used in the Keeler 
Lithium Chargers. Do not try to insert a Keeler handle with a 
blue base into the Keeler Lithium Chargers. Refer to Keeler 
handle and bulb identification. 

Содержание Ophthalmoscope

Страница 1: ...Next Next Ophthalmic and Diagnostic Instruments Ophthalmoscope Retinoscope Otoscope Instructions for use...

Страница 2: ...y conditioning charging 21 7 GenMed Wall Unit 22 Wall mounting Wall mounting template 23 Power supply assembly 24 Connecting securing heads 25 Using the GenMed Wall Unit 26 Dispos A Spec 8 Lithium Cha...

Страница 3: ...ons and other information contained in this document without prior notice The products listed opposite are registered trademarks of Keeler Ltd 2012 Copyright Keeler Limited 2012 Published in the UK 20...

Страница 4: ...in the design development and manufacture of this product to ensure that you get many years of trouble free service However it is important that you read the descriptions installation and operating i...

Страница 5: ...on the product or on its packaging and instructions indicates that it was put on the market place after August 2005 and that this product shall not be treated as Household Waste Type B protections aga...

Страница 6: ...odically inspect for signs of damage Do not use in the presence of flammable gases liquids or in an oxygen rich environment This product should not be immersed in fluids Do not position the device so...

Страница 7: ...nce on cleaning routine maintenance to prevent personal injury damage to equipment Failure to carry out recommended routine maintenance as per the instructions in this IFU may reduce the operational l...

Страница 8: ...es Do not use Keeler Retinoscopes in ambient temperatures above 30 C Disposable speculae should not be used for insufflation testing Plastic reusable Speculae will degrade if exposed to ultra violet l...

Страница 9: ...battery contacts to avoid short circuiting during disposal After removal of the battery do not touch the battery contacts and the patient simultaneously Note Lithium Ion batteries contain no toxic he...

Страница 10: ...l procedures While no visible retinal lesions have been identified for ophthalmic instruments it is recommended that illumination levels be set to the minimum level necessary to perform the diagnostic...

Страница 11: ...ofessional Specialist Pocket Auxiliary Lens Wheel Lens Wheel Pupilometer Lens Wheel Graticule and Filter Control Lens Wheel Auxiliary Lens Wheel Filter Control Graticule Control 5 Instrument heads Oph...

Страница 12: ...tioner and Standard 40D to 25D Pocket 20D to 20D Graticule Control The graticule control is used to select the required beam for examination The choice of graticules is as follows Wide Angle Illuminat...

Страница 13: ...andard Pocket Filter Control The filter control is used to select the required filter Professional Practitioner Standard only Filter Applications Red Free Green filter Is used to examine the blood ves...

Страница 14: ...med between the mid position and the bottom position The focussing and axis control can be rotated continuously in any direction Focussing and axis control Spot The vergence is altered by sliding the...

Страница 15: ...oscopes Five permanent Speculae are provided with each Otoscope Set The diameters are as follows 2 5 3 5 4 5 5 5 8mm These are attached to the Otoscope head as shown in the following diagrams Standard...

Страница 16: ...is An Insufflation adaptor is also available for Practitioner as shown on the previous page Minor surgical Procedures Should you wish to use surgical instruments such for minor procedures the followin...

Страница 17: ...voltage See base of bulb Blue 2 8v for dry cell battery handles Red 3 6v for rechargeable handles Black LED Loosen the set screw securing the instrument head to the handle GenMed Wall Unit only Remov...

Страница 18: ...Next Home Back 18 6 Instrument handles 2x AA or MN1500 Blue Cap Rechargeable Battery Red Cap 2 x AA or MN1500 Blue Cap Slimline Pocket C size Rechargeable Battery Red Cap Blue Cap 2 x C or MN 1500...

Страница 19: ...st Professional Standard and Practitioner Ophthalmoscopes and Keeler Retinoscopes are compatible with Keeler 2 8v and 3 6v Keeler handles The Keeler LED module is only compatible with Keeler 2 8V and...

Страница 20: ...ies Unscrew battery cap insert batteries and replace battery cap as shown on page 18 Please note Keeler rechargeable handles are normally supplied complete with a rechargeable battery 3 6v 0 7Ah Ni Cd...

Страница 21: ...RECHARGING UNTIL THE BATTERY IS COMPLETELY EMPTY Step 3 Once empty recharge the battery until full This will take approximately 15 hours Repeat steps 1 2 and 3 three times i e fully charge and discha...

Страница 22: ...intended mounting position For the Gen Med Wall Units drill two holes 6mm x 55mm deep and 110mm apart For the Dispenser Unit drill an additional two holes 249mm below the existing holes using the tem...

Страница 23: ...Next Home Back 23 7 GenMed Wall Unit and Dispenser mounting template THIS WAY UP WALL UNIT AND DISPENSER TEMPLATE Scale 1 1 Do not scale 249mm 110mm...

Страница 24: ...ed Wall Unit power supply assembly Set Plug Replace the blanking plate with the appropriate mains plug adaptor if required or use IEC 60320 TYPE 7 connector not supplied IEC 60320 TYPE 7 connector Con...

Страница 25: ...ted by the GenMed Wall Unit If such effects are suspected switch off the offending equipment Connecting your instrument head to the wall unit handle The instrument head should be screwed positively on...

Страница 26: ...returned correctly to its cradle and that the LED goes out Only one handle can be used at a time Replace the handle before using the other instrument Refer to the instructions in section 5 for informa...

Страница 27: ...charge Solid LED Battery is charging The handle can be used at any time during the charging cycle and will automatically resume charging when handle is placed back in the charging well When using the...

Страница 28: ...hods of sterilization must not be used The cleaning and sterilization of reusable speculum metal tongue depressor nasal dilator laryngeal and post natal mirrors can be accomplished as follows a Manual...

Страница 29: ...2 5amps Operation Maximum 15 minutes ON Minimum 10 minutes OFF Classification Class II equipment Type B protection against shock Instrument heads and handles Input voltage DC 3V 2xAA Alkaline Batterie...

Страница 30: ...ese instruments When installing or using these instruments please read carefully and observe what is described here Portable or mobile type radio frequency communication units may have an adverse effe...

Страница 31: ...t only Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short in...

Страница 32: ...the RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Keeler chargers and power systems are suitable for use in all e...

Страница 33: ...d strengths from fixed RF transmitters as determined by an electromagnetic site survey2 should be less than the compliance level in each frequency range 3 Interference may occur in the vicinity of equ...

Страница 34: ...ommunications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80MHz d 1 2 p 80MHz to 800MHz d 1 2 p 800MHz to 2 5GHz d 2 3...

Страница 35: ...x Professional Streak retinoscope LED 1015 P 7023 2x 3 6v Halogen Standard Deluxe otoscope bulbs 1015 P 7031 2x 2 8v Halogen Standard Deluxe otoscope bulbs 1015 P 7058 2x 3 6v Practitioner Fibre Optic...

Страница 36: ...authorised Keeler representatives Your Keeler product is guaranteed for 3 years and will be replaced or repaired free of charge subject to the following Any fault due to faulty manufacture The instru...

Страница 37: ...r European Countries with separate Collection Systems This symbol on the product or on its packaging and instructions indicates that it was put on the market place after August 2005 and that this prod...

Отзывы: