KEELER LOUPE FRAME
2. CLEANING AND DISINFECTION INSTRUCTIONS
Only manual non-immersion cleaning as described should be used for this instrument. Do not
autoclave or immerse in cleaning fluids.
1. Wipe the external surface with a clean absorbent, non-shedding cloth dampened with
a de-ionised water / detergent solution (2% detergent by volume) or water / isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces.
2. Ensure that excess solution does not enter the instrument. Use caution to ensure cloth is
not saturated with solution.
3. Surfaces must be carefully hand-dried using a clean non-shedding cloth.
4. Safely dispose of used cleaning materials.
3. SETTING UP AND USING THE LOUPE FRAME
1. Fit the Loupe unit to the frame via the pin mounting system as shown.
2. The protective shield is fitted by pushing the Shield into position as shown.
Protective shield in correct position