20
EN
ALL PUPIL II by KEELER
Immunity test IEC 60601 Test
level
Compliance
level
Electromagnetic
environment – guidance
Electrical fast
transient/burst.
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for power
supply lines
± 2 kV for power
supply lines
N/A
Mains power quality should be
that of a typical professional
healthcare facility environment.
Surge.
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) for
input/output line(s)
± 1 kV line(s) to
line(s)
N/A
Mains power quality should be
that of a typical professional
healthcare facility environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
U
T
= 0% 0.5 cycle
(0, 45, 90, 135, 180,
225, 270, 315°)
U
T
= 0%; 1 cycle
U
T
= 70%;
25/30 cycles (@ 0°)
U
T
= 0%;
250/300 cycle
U
T
= 0% 0.5 cycle
(0, 45, 90, 135, 180,
225, 270, 315°)
U
T
= 0%; 1 cycle
U
T
= 70%;
25/30 cycles (@ 0°)
U
T
= 0%;
250/300 cycle
Mains power quality should be
that of a typical
professional
healthcare facility environment.
If the user of the Keeler All Pupil
II requires continued operation
during power mains interruptions,
it is recommended that the
charger be powered from an
uninterruptible power supply.
Power frequency
(50/60 Hz)
Magnetic field. IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at a level characteristic
of a typical location in a typical
professional healthcare facility
environment.
Note:
U
T
is the a. c. mains voltage prior to application of the test level.
Immunity
test
IEC 60601
Test level
Compliance
level
Electromagnetic environment
– guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Keeler
All Pupil II, including cables, than the recommended
separation distances calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Conducted
RF IEC
61000-4-6
6 Vrms
150 kHz to 80
MHz
6 V
d = 1.2 √ p
Radiated
RF IEC
61000-4-3
10 V/m 80MHz
to 2.7GHz
10 V/m
d = 1.2 √ p 80MHz to 800 MHz
d = 2.3 √ p 800MHz to 2.7GHz