KD KDS-1013 Скачать руководство пользователя страница 2

 

Display: 0.5secretary 

7)  Bar  graph:50dB  scale  at  1dB  step  for  monitoring  current  sound  pressure  level  display 

period:50mS 

8) level ranges: 30-80dB; 50-100dB; 

      60-110dB; 80-130dB 

9) Over indicate over range: 

      Under indicate less than lower limit of the range. 

10) Ac output: 0.707Vrms at FS output impedance 

approx 600

Ω 

11) DC output: 10mV/dB output impedance approx 100

Ω 

12) weighting time : fast/slow 

13) Microphone :1/2 inch electret condenser microphone 

14) Max: Maximum hold 

15) Power supply: 1*9V alkaline cells or DC 9V adapter   

Power life: about 30hrs(alkaline cells) 

16) self calibration time: 10 sec(every turn on) 

17) Operating Temperature: 0

~40

 

Operating Humidity: 10%~80%RH 

18) Storage Temperature:-10

~60

 

Storage Humidity :10%~70%RH 

19) Dimensions: 265(L)*72(W)*35(H)mm 

20) Weight: 300g(including batteries) 

 

D. Calibration Procedures: 

Please use a Standard Acoustic Calibrator. 

1) Make the following settings: 

Display: SPL(dBA) time 

weighting: FAST 

Level range: 60~110dB 

Measurement mode: MAX function disable 

2) Insert the microphone carefully into the 1/2 inch hole of the Calibrator. 

3) Turn on the Calibrator and adjust the Protentiometer in side the Battery compartment of the 

unit (shown in the diagram 1).    The level display will indicate the desired Level (94.0). 

Our products are all well calibrated before Shipment. Recommended recalibration cycle: 1 year 

 

 

Отзывы: