RATE can be selected from the main menu or will be displayed automatically after the 
Volume setting. 

            
 

The display reads: 

Rate:  X.X ul/h  or  X.X ml/h 

 
 

Rate units are ul/h  or ml/h, microliter or milliliter/hour and are selected automatically 
from the syringe diameter. 

            

1)

 

Use the 

arrow 

keys to display the flowrate required. 

2) 

Enter the rate with the 

select

 key. 

3) 

If the rate entered is out of range, a message is displayed: 

 

 " Max Rate XX ml/h"

    

 

To continue enter a lower rate ,or select a larger syringe. 

 

note:

 

a. 

If a dispense volume was set then the display will now change to  

 

 

 

 

"Volume : 0.0" ,that is, in automatic dispense mode. 

 
 

When the pump runs the actual dispensed volume will be displayed and will 

 

increment until the set volume is reached, at which time the pump will stop 

 

automatically. 

 
 
 

b. 

If no dispense volume is set the LCD displays the set flowrate and,  

 

 

 

when running, the directional indicator arrow will pulse.   

 

  

START THE PUMP

 

 

a.   Press the 

run/stop 

key to start the pump.  A second press will stop the pump. 

 
b.   If the pump is stopped during a dispense the volume accumulator is not cleared - the 

dispense is paused.  Restarting the pump with a second press of the 

run/stop

 key 

continues the dispense to the target volume. 

 

POWER FAILURE MODE 

 
If there is a temporary power interruption, when no dispense volume is selected, the 

Power Up 

run/stop

 option allows the pump to either resume the dispense when the power returns, or 

remain stopped but display a message warning of the power failure. 
 
1)   Move the main menu to the right and select 

PWR UP

 
2)   The display will read 

Power Up  Run 

( or STOP ).  Use the 

arrow 

keys to change display       

       and 

select

 the required option 

 

Run 

After power is returned the pump resumes normal operation.  The display flashes

 

PowerFailed 

to acknowledge a power interruption.  Press Select to display the Rate. 

 
STOP 

Pump does not run when power returns.  The LCD displays the rate setting.  

 
If a dispense volume is entered the power interruption stops the pump, clears the volume 
accumulator and returns the volume to the setpoint.  Only if the power failure  setting is “run” will 
the power failure warning be displayed.  
 

CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING 

 
While the pump is running it is possible to look at the volume and flowrate settings without 
interrupting the operation of the pump.  If required the volume and flowrate settings can be 

Содержание KDS100

Страница 1: ...kd Scientific Model 100 Series ...

Страница 2: ...ge diameter entry 5 volume entry 5 rate setting 6 power failure mode 6 rate and volume changes while running 7 fast forward 7 Syringe loading 8 Stall settings microliter syringe precautions 9 NV Ram Failure 10 Fuses 10 Maintenance 10 Syringe Table 11 Table of Maximum and Minimum Flow rates 12 Warranty 13 ...

Страница 3: ... 0 00106 cm hr maximum 12 2 cm hr Minimum stepping rate one step in thirty seconds may vary depending on syringe size Maximum stepping rate 400 steps second Speed range 12000 1 Flowrate range 0 1ul hr 10 ul syringe to 506ml hr 60 ml syringe Linear force 20 lb min 9 Kgm Dimensions 9x6x5 inch 23x15 25x13 cm Weight 5 lb 2 Kgm Atmospheric Specifications Temperature 40 to 85 Degrees C Humidity L T 95 M...

Страница 4: ...ectly or the syringe can be identified from a table of syringes held in memory When the syringe is selected from the table the diameter is entered automatically Two dispense modes are available a dispense volume mode in which the pump keeps track of the volume dispensed and automatically stops the pump when a set target volume is reached b run mode where the pump runs at the set flowrate until man...

Страница 5: ...t the right rear corner of the pump When the power is turned on the LCD will display the Volume or Rate setting Press the select key to return to the main menu MAIN MENU Pressing select repeatedly will always return the display to the main menu The main menu consists of five variables three of which are displayed at all times The center variable pulses to indicate that this option can be reviewed ...

Страница 6: ...rs the internal syringe diameter See Table 1 for the syringe library b Direct entry From the main menu select DIA The display will read xx xx mm Use the arrow keys to display the measured internal diameter of the syringe in millimeters and enter with the select key Changing the diameter clears the Volume and Rate settings and the display will now prompt for the new Volume setting VOLUME ENTRY The ...

Страница 7: ...t cleared the dispense is paused Restarting the pump with a second press of the run stop key continues the dispense to the target volume POWER FAILURE MODE If there is a temporary power interruption when no dispense volume is selected the Power Up run stop option allows the pump to either resume the dispense when the power returns or remain stopped but display a message warning of the power failur...

Страница 8: ...ate and the volume display continues to increment uninterrupted to the new dispense setting note If the VOLUME is changed to a setting smaller than the volume already accumulated then the pump stops when the new smaller dispense volume is selected CHANGE OR REVIEW RATE SETTING WHILE RUNNING 1 Press select to change the volume accumulator display to the main menu 2 Select RATE 3 Make rate changes i...

Страница 9: ...yringe holder making sure that the flange of the syringe barrel is pressed against the side of the syringe holder Rotate and release the syringe clamp to hold the syringe in place Press in button 2 and move the pusher block back along the guide rods to make contact with the syringe plunger GLASS SYRINGES With some glass syringes the corners of the flange of the syringe barrel are rounded and can c...

Страница 10: ...he locking screw on the Adjustable syringe rest and position the rest so that the plunger of the loaded syringe is pressed firmly against the pusher block and the syringe barrel flange is pressed up to the syringe rest The barrel of the loaded infusion syringe should be clamped in the V of the syringe holder with the spring loaded clamp Tighten the screw on the syringe rest to lock the rest in pla...

Страница 11: ...f clearing the NV Ram Failure is to press select and then enter a changed rate setting 7 FUSES The fuses are located in the power entry module on the rear panel The linecord must be first removed to gain access to the fuse holder FUSE 5x20mm 250V slow blow US model 0 1AT slow blow CE model 0 05AT slow blow 8 MAINTENANCE Maintenance is required only for the moving mechanical parts which should be k...

Страница 12: ...3 Becton Dickinson 8 Sherwood Glass all types Monojet Plastic 0 5cc 4 64mm 1cc 4 65mm 1 4 64 3 8 94 2 5 8 66 6 12 70 5 11 86 12 15 90 10 14 34 20 20 40 20 19 13 35 23 80 30 22 70 50l 26 60 60 28 60 4 Hamilton 9 Terumo 1000 Series Gastight 10ul 0 48mm 1cc 4 73mm 25 0 78 3 9 00 50 1 03 5 13 04 100 1 46 10 15 79 250 2 30 20 20 18 500 3 26 30 23 36 60 29 45 1ml 4 61mm 2 5 7 28 5 10 30 10 14 57 25 23 0...

Страница 13: ...ml h 219ml h 30ml 0 1ml h 282ml h 60ml 0 2ml h 426ml h Model 101 Syringe size Minimum Maximum 10ul 0 001ul m 0 382ul m 25ul 0 001ul m 1 010ul m 50ul 0 001ul m 1 762ul m 100ul 0 001 ul m 3 542 ul m 250ul 0 01 ul m 8 78 ul m 500ul 0 01ul m 17 65ul m 1ml 0 1 ul m 35 2 ul m 2 5ml 0 1 ul m 88 0 ul m 3ml 0 1 ul m 122 5 ul m 5ml 0 1 ul m 176 2 ul m 10ml 0 001 ml m 0 351 ml m 20ml 0 001 ml m 0 602 ml m 30...

Страница 14: ...ied warranties of merchantability or fitness for a particular use or of any other nature Some States do not allow this limitation on implied warranty so the above limitation may not apply to you If a defect arises within the warranty period contact KD Scientific Inc 3307 Windy Bush Road New Hope PA 18938 tel 215 862 6477 fax 215 862 6478 The customer is responsible for shipping charges and must fi...