Battery Charging Instructions

1.  Plug the AC power cord into the DC power 

supply.

2.  Plug the AC wall plug into an AC wall outlet.

3.  Locate the arrow on the charging cord 

connector.

4.  Place therapy unit and charging cord so that 

touch screen (on therapy unit) and arrow (on 
charging cord connector) both face up.

5.  Plug charging cord connector 

securely

 into 

therapy unit.

The plug indicator appears on the touch 
screen while the unit is plugged into a 
wall outlet.

When the unit is correctly plugged into the power 
supply, the Battery Charging Indicator Light will 
glow amber as the battery charges. When the 
battery has reached full charge the light will glow 
green.

Battery Charge Level Indicator

The battery charge level is shown on the bottom of 
the touch screen.

Fully Charged

In Use

Battery Low (charge battery soon)

Battery Critical (charge battery 
immediately)

It should take approximately six hours 
to fully charge the battery. To maximize 
battery life, keep the unit plugged in 
whenever possible.

 Battery  Charging

Indicator Light

Plug to DC

Power Supply

Battery Level Indicator

AC 

Power 

Cord

AC Wall Plug

Charging

Cord 

Connector

White

Arrow

Canister Parts

Canister latch guide on the therapy unit 
may have sharp edges. Do not handle 
the ActiV.A.C.

®

 Therapy Unit by the 

canister latch guide.

When not in use, always store the 
ActiV.A.C.

®

 Therapy Unit in the carrying 

case without a canister in place.

Always apply canister straight on and 
straight off the therapy unit. Do not 
twist or turn canister when installing or 
removing.

Canister Changes

In the US, additional dressings and 
canisters may be ordered by calling KCI 
at 1-800-275-4524. Please allow at least 
3-5 business days for delivery. Outside 
the US, please contact your local KCI 
representative.

The ActiV.A.C.

®

 canister should be 

changed when full (the alarm will 
sound), or at least once a week to 
control odor.

1.  Stop therapy by pressing 

Therapy 

On/Off

 on the touch screen. 

Do not 

turn power off to the therapy 
unit.

2.  Slide both tubing clamps toward the tubing 

connector.

3.  Close both clamps tightly. Several

clicks should be heard.

Canister 

Latch 

Release

Canister 

Latch 

Guide

On/Off

4.  Compress and then twist to

separate tubing connectors.

5.  Press the canister latch release.

6.  Remove canister from 

therapy unit.

Dispose of the canister according to 
institution and/or local environmental 
regulations.

7.  Install the new canister onto the therapy unit. An 

audible click should be heard when the canister 
is properly installed.

8.  Ensure the canister is securely installed on the 

therapy unit. If the canister is properly installed, 
it cannot be removed by gently pulling it directly 
away from the unit.

9.  Compress and then twist to 

reconnect dressing tubing
connectors.

10. Open both tubing clamps.

11.  Press the 

Therapy On/Off

 button on 

the touch screen to restart therapy. 
Verify the dressing collapses

.

On/Off

Seal Check™ Leak Detector

When the ActiV.A.C.

®

 Therapy Unit detects a 

signifi cant leak, the 

Leak Alarm

 will activate. (See 

Alerts and Alarms

 section of the user manual for 

more details on this alarm.)

Press this button to use the Seal Check™ 
Leak Detector to help fi nd leaks.

The Seal Check™ feature uses an audible tone and 
bar graph to help fi nd leaks. The frequency of the 
audible tone and the height of the bar graph will 
refl ect the leak rate. The audible tone slows down 
and the bar graph decreases in height as the leak 
is found. Orange bar graph indicates a signifi cant 
leak. Green bar graph indicates that the ActiV.A.C.

®

 

Therapy System is operating normally.

Most leaks occur:

• 

where the drape meets the skin. 

• 

where the SensaT.R.A.C.

®

 Pad  is attached 

to the drape. 

• 

at tubing connections.

• 

when the canister is not securely 
connected to the therapy unit.

1.  Ensure connector between dressing tubing and 

canister tubing is properly locked.

2.  Ensure the canister is securely installed onto the 

therapy unit.  

3.  While therapy is on and using light pressure, 

move your hand and fi ngers slowly around the 
edges of the drape and SensaT.R.A.C.

®

 Pad. The 

bar graph will lower and the frequency of the 
audible tone (if Seal Audio is on) will decrease 
when the leak is found.

4.  Refer to the 

Application Instructions

 provided 

with V.A.C.

®

 Dressings for information on using 

excess V.A.C.

®

 Drape material to seal the leak 

area.

5.  If necessary, contact your doctor or nurse or KCI 

for assistance or further information.

Press ‘Seal Check

’ or ‘?’ for more information

Audio Pause

Reset

Seal

Check

TM

Leak Alarm

Flashing 
Green Oval

Seal

Check

TM

Seal Check

Low 

High 

Leak Rate

Exit

Seal Audio

Line on bar 

graph is the 

transition 

point from 

green to 

orange and 

vice-versa.

Press to turn 

the Seal 

Audio tone 

on or off.

Отзывы: