Chapter 7. Maintenance and Troubleshooting
40
Kaye LTR -90 Operating Manual
7.2.4 Change the Fuse
The Calibrator has a product fuse that protects from overcurrent.
WARNING! To prevent possible electrical shock, fire, or personal injury, use only specified replacement
parts.
To replace the fuses (see Figure 24 below):
1.
Disconnect the mains-power cord from the power-entry module.
2.
Open the power entry module and remove the fuse holder.
3.
Replace the fuse with exact replacements as indicated in the specifications in Chapter 2.
Figure 24: Changing the Product Fuse