Chapter 6. Calibrator Operation
6.1.4 Set a Soft Cutout
Soft Cutout is an upper-heat limit that protects the probes. The Hard Cutout protects the Calibrator and it cannot be
changed (set by factory).
To set a Soft Cutout (see Figure 16 below):
1.
Push .
2.
Push .
3.
Push .
4.
Push
to select the Soft Cutout.
5.
Push
and
to highlight a digit, then push
to increase the digit or push
to decrease the digit..
6.
Push
to save the Soft Cutout.
Figure 16: Soft Cutout
6.1.5 Reset an Over-Temperature Cutout
If the Soft or Hard Cutout is exceeded, the Calibrator enters Cutout Mode and must be reset.
To exit Cutout mode and reset the Calibrator:
1.
Let the Calibrator cool to below the Soft Cutout limit.
2.
After it has cooled, push
followed by
to reset the Calibrator.
66.03°C
EDIT PRESETS
PRESET 1:
PRESET 2:
PRESET 3:
PRESET 4:
+0080.00
°C
+45.00
°C
+26.00
°C
-26.00
°C
Содержание LTR -90
Страница 4: ...ii no content intended for this page...
Страница 14: ...Chapter 2 Specifications 8 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 24: ...Chapter 4 Parts and Controls 18 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 27: ...Kaye LTR 90 Operating Manual 21 Chapter 5 General Operation Figure 7 Insert Installation and Removal...
Страница 29: ...Kaye LTR 90 Operating Manual 23 Chapter 5 General Operation Figure 8 Probe Insertion Figure 9 Rubber Insulator...
Страница 30: ...Chapter 5 General Operation 24 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 44: ...Chapter 6 Calibrator Operation 38 Kaye LTR 90 Operating Manual no content intended for this page...
Страница 49: ......