In order to prevent negative effects on the medical device, make
sure that the interior and the exterior of the medical device are dry
after completion of the cycle, and then grease it immediately with
servicing agents from the KaVo care system.
7.5 Drying
Manual Drying
▶
Blow off the outside and inside with compressed air until water drops
are no longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the ther‐
modisinfector.
Reprocessing steps in accordance with ISO 17664
45