Select a suitable procedure (depending on the available autoclave) from
the following sterilisation processes:
▪ Autoclave with pre-vacuum:
–
at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F)
Drying time: 20 min.
▪ Autoclave using the gravity method:
–
at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F) or alternatively
Drying time: 30 min.
–
at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4
°F)
Drying time: 30 min.
▶
Use according to the manufacturer's Instructions for Use.
Preparation methods according to ISO 17664
55