KaVo SMARTtorque LUX S619 L Скачать руководство пользователя страница 16

7.3.1

   

Manual external disinfection

KaVo recommends the following products based on material compatibility.

▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)

▪ FD 322 made by Dürr

▪ CaviCide made by Metrex

Consumables required:

▪ Cloths for wiping off the medical device.

▶ Spray the disinfectant on a cloth, then thoroughly wipe down the medical de‐

vice and leave the disinfectant to soak in according to the instructions from the

disinfectant manufacturer.

▶ Follow the instructions for use of the disinfectant.

7.3.2

   

Manual disinfection - internal

The efficacy of manual internal disinfection must be demonstrated by the manu‐

facturer of the disinfection agent. With KaVo products, use only disinfection

agents that have been released by KaVo with respect to the compatibility of mate‐

rials (e.g. WL-cid / made by ALPRO).

▶ Immediately after internal disinfection, lubricate the KaVo medical device im‐

mediately with care agents from the KaVo care system.

▶ Follow the instructions for use of the disinfectant.

7.3.3

   

Machine disinfection - external and internal

KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which

are operated with alkaline cleaning agents with a pH value of max. 10 (e.g. Miele

G 7781 / G 7881 – Validation was carried out with Programme "VARIO-TD",

cleaning agent "neodisher

®

 mediclean", neutralisation agent "neodisher

®

 Z" and

rinsing agent "neodisher

®

 mielclear" and only applies to the material compatibility

with KaVo products).

▶ For program settings as well as cleansers and disinfectants to be used, please

refer to the Instructions for Use of the thermodisinfector (complying with max.

pH value of 10).

▶ In order to prevent negative effects on the medical device, make sure that the

interior and the exterior of the medical device are dry after completion of the

cycle, and then grease it immediately with servicing agents from the KaVo

care system.

 

 

 

 

 

 

 

 

 

 

 

Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643

SMARTtorque S619 C - 1.008.1642
7 Reprocessing methods according to ISO 17664 | 7.3 Disinfection

14/21

Содержание SMARTtorque LUX S619 L

Страница 1: ...Instructions for use SMARTtorque LUX S619 L 1 008 1641 SMARTtorque Mini LUX S615 L 1 008 1643 SMARTtorque S619 C 1 008 1642 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...the supply hose 11 6 2 Cleaning the spray nozzle 11 7 Reprocessing methods according to ISO 17664 12 7 1 Preparations at the site of use 12 7 2 Cleaning 12 7 2 1 Manual external cleaning 12 7 2 2 Aut...

Страница 4: ...e technicians Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable CE mark European Community A product bearing this mark meets the requirements of the applicable EC directiv...

Страница 5: ...can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a sit...

Страница 6: ...adle without the cutter or grinder CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Reduced product life The medical device should be cleaned serviced and...

Страница 7: ...ant national statutory regulations Proper use According to these regulations this medical device may only be used for the de scribed application by a knowledgeable user The following must be observed...

Страница 8: ...ion and storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very...

Страница 9: ...ount the MULTIflex LUX MULTIflex LED coupling Screw the MULTIflex LUX MULTIflex LED coupling onto the turbine hose and tighten with the wrench Rotate the spray ring on the MULTIflex LUX MULTIflex LED...

Страница 10: ...and the medical device and check the following pressures Drive pressure 2 8 bar 40 psi Return air 0 5 bar 7 psi Water 0 8 to 2 5 bar 12 to 36 psi 4 4 Checking the O rings CAUTION Missing or damaged O...

Страница 11: ...the medical device is securely locked onto the MULTIflex LUX MULTIflex LED coupling by pulling on it Mount the medical device accurately on the MULTIflex LUX MULTIflex LED coupling and push it backwa...

Страница 12: ...ng and remov ing use gloves or a fingerstall to prevent an injury or infection Forcefully press the push button with your thumb and simultaneously insert the cutter or grinder all the way Check that t...

Страница 13: ...front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzle CAUTION Hazard from insufficient cooling water Overheating of the medical device and damage to the tooth Check th...

Страница 14: ...d Tap water 30 C 5 C 86 F 10 F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Automated external cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1...

Страница 15: ...dance with EN ISO 15883 1 7 2 4 Automated internal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max...

Страница 16: ...Vo care system Follow the instructions for use of the disinfectant 7 3 3 Machine disinfection external and internal KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are opera...

Страница 17: ...er servicing and care Reduced product life Perform regular proper care and maintenance Note KaVo only guarantees that its products will function properly when the care prod ucts used are those listed...

Страница 18: ...ROcare PLUS See also Instructions for Use KaVo QUATTROcare PLUS 2124 A Servicing of the clamping chuck KaVo recommends cleaning and servicing the chuck system once a week See also Instructions for Use...

Страница 19: ...with EN 13060 ISO 17665 1 CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Before each sterilisation cycle service the medical device with KaVo care produ...

Страница 20: ...ted from dust with minimum ex posure to germs in a dry dark and cool space Comply with the expiry date of the sterilised items Instructions for use SMARTtorque LUX S619 L 1 008 1641 SMARTtorque Mini L...

Страница 21: ...head INTRA 0 411 9911 Cleanpac 10 units 0 411 9691 Material summary Mat no Adaptor INTRAmatic CLEANspray and DRYspray 1 007 1776 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo...

Страница 22: ...ually extend to lamps light conductors made of glass and glass fibres glassware rubber parts and the col ourfastness of plastic parts All liability is excluded if defects or their consequences origina...

Страница 23: ...Phone 420 272 654 566 Fax 420 272 654 732 Email info kavo cz Internet www kavo cz Russia KaVo Dental Russland GmbH nab reki Fontanka 130 A 190005 St Petersburg Phone 7 812 331 86 96 Fax 7 812 251 06...

Страница 24: ......

Страница 25: ...1 008 1927 kb 20111124 03 en...

Отзывы: