Instructions for use SMARTair Evo, SMARTair Evo mobile, SMARTair Evo FLEXspace
6 Operation | 6.5 Setting the response sensitivity
36 / 46
3
2
1
▶
Press the power switch to switch the suction device on.
▶
Press the "Automatic" button
③
(for approx. 3 seconds), until LEDs
①
and
②
flash.
ð
The suction device is in "Setting mode".
Setting and saving the response sensitivity
▶
Turn on the power switch on the control unit.
▶
Start the working device at a speed of 5000 to 7000 min
-1
and keep it run-
ning while performing the setting.
Case 1: LEDs do not light up
▶
Briefly press the "Plus"
⑤
button several times until all LEDs
②
light up.
▶
Press the "Automatic"
③
button to save the setting.
Case 2: All LEDs light up
▶
Briefly press the "Minus"
④
button several times until all LEDs
②
go off.
▶
Briefly press the "Plus"
⑤
button several times until all LEDs
②
light up.
▶
Press the "Automatic"
③
button.
ð
The values are saved.
ð
The unit is in "normal mode" again.