4 PC requirements
29
-
Do not position the PC where it could be splashed
with liquids.
-
Clean the PC in accordance with the
manufacturer’s instructions.
-
The PC to be used with the unit must be installed
in a location that meets all local and national safety
requirements with regards the connection of a PC
to an x-ray device.
-
The connection of the unit to the PC must meet IEC
60601-1 requirements.
-
Connecting the unit to an IT-network that includes
other equipment or changing the IT-network can
cause unidentified risks to patients or operators. It
is the responsibility of the organization controlling
the IT-network to identify, analyse, evaluate and
control these possible risks.
■
The use of accessory equipment not complying
with the equivalent safety requirements of this
equipment may lead to a reduced level of safety of
the resulting system.
Consideration relating to the choice shall include:
-
Use of the accessory in the PATIENT VICINITY.
-
The safety certification of the accessory has been
performed in accordance to the appropriate IEC
601-1 and/or IEC 601-1-1 harmonized national
standard.
4.2 The dental imaging software
The dental imaging software installed in the PC that is
used with the unit must have CE-mark according to
Medical Device Directive, for example CLINIVIEW™
software.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...