8 Special procedures
138
sions.txt included in the software release package. If all
versions match, proceed to the next step. Otherwise,
you must first update the mismatching components.
2.
Set the software release version by pressing F9, or by
manually typing:
pmode 0 <ver_major> <ver_minor>
Example:
pmode 0 2 00
The unit should reply
programming ended successfully, version now
R2.00
This command makes the unit store the versions of the
software components currently loaded, so that the va-
lidity of the software release may be verified later. If one
or more software component are manually changed
without this step being carried out, the version will be
displayed with an "M" postfix (as in "Modified"), e.g.
R1.11M.
8.8 Volatile installation
Volatile installation is useful when a circuit board is not
responsive to Ethernet communication, for example due to
corrupted software. Using a JTAG programmer the
software may then be loaded directly into the RAM of the
circuit board, after which the software may be permanently
installed the normal way through Ethernet. Volatile
installation is available for all embedded software
components except the GUI.
To perform a volatile installation you will need:
An Altera USB-Blaster Download Cable, available from
Altera at:
http://www.altera.com/products/devkits/kit-cables.html
Alternatively, a Terasic USB Blaster Download Cable,
available from Terasic can be used:
www.terasic.com.tw
Quartus II Programmer software, available for download
from Altera at:
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X Ray System Installation Manual ENGLISH 216363 rev 1 0 805 5080...
Страница 2: ......
Страница 4: ......
Страница 8: ...iv rev...
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Страница 29: ...2 Unit description 21...
Страница 30: ...2 Unit description 22...
Страница 38: ...4 PC requirements 30...
Страница 54: ...5 Installing the unit 46 4 Fasten the covers...
Страница 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Страница 104: ...7 Cephalometric calibration and alignment 96...
Страница 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Страница 117: ...7 Cephalometric calibration and alignment 109...
Страница 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Страница 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Страница 158: ...8 Special procedures 150...
Страница 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Страница 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Страница 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...
Страница 168: ...9 OP 3D Pro Pre sales check list 160...