6 Troubleshooting | 6.3 Changing the water filter
CAUTION
Hazard from insufficient spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check the filter and exchange if necessary.
▶
Check the spray water channels and clean the spray nozzles with the nozzle
needle
Mat. no. 0.410.0921
if necessary.
▶
Unscrew the sleeve
①
in counterclockwise direction from the insert
②
and
pull it off.
▶
Unscrew the water filter
③
with the wrench (Mat. no. 1.002.0321) and take
it out
▶
Insert the new filter (Mat. no. 1.002.0271) and screw it in with the wrench
▶
Place the sleeve
①
on the insert
②
, and screw it tight in clockwise direction.
17 / 30