7.2.2 Automated external cleaning
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
which are operated with alkaline cleaning agents with a pH value of max.
10 (e.g. Miele G 7781/G 7881 – Validation was carried out with pro‐
gramme "VARIO-TD", cleaning agent "neodisher® mediclean", neutralisa‐
tion agent "neodisher® Z" and rinsing agent "neodisher® mielclear" and
only applies to the compatibility of materials with KaVo products).
▶ For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector
(complying with max. pH value of 10).
▶ In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry after
completion of the cycle, and then grease it immediately with servicing
agents from the KaVo care system.
7 Preparation methods according to ISO 17664