Instructions for use MASTERtorque LUX M9000 L COLOR - 1.007.9900 MASTERtorque LUX M9000 L COLOR -
1.007.7100
6 Troubleshooting | 6.1 Changing O-rings on the MULTIflex (LUX) / MULTIflex LED coupling
16 / 30
6 Troubleshooting
6.1 Changing O-rings on the MULTIflex (LUX) /
MULTIflex LED coupling
NOTICE
Improper care of the O-rings.
Malfunction or complete failure.
▶
Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wetted
with KaVo Spray.
▶
Press the O-ring between your fingers to form a loop.
▶
Push the O-ring to the front, and remove it.
▶
Insert new O-rings into the grooves.
6.2 Cleaning the spray nozzle
WARNING
Hazard from nonsterile products.
Infection danger to the care provider and patient.
▶
Reprocess and sterilise the medical device properly before the next use.
CAUTION
Hazard from insufficient spray water.
Overheating of the medical device and damage to the tooth.
▶
Check the spray water channels and clean the spray nozzles with the nozzle
needle
Mat. no. 0.410.0921
if necessary.
▶
Check the water filter and exchange if necessary.
6.3 Changing the water filter
WARNING
Hazard from nonsterile products.
Infection danger to the care provider and patient.
▶
Reprocess and sterilise the medical device properly before the next use.