Instructions for use MASTERspace® - Workplace
3 Product description | 3.3 Technical data
Electrical connection
Rated voltages
100 V or 110 - 120 V or 230 V at 50
- 60 Hz
Power consumption
max. 16 A
Connected load at the multiple
socket outlet
(for UL versions)
max. 12 A
Leakage current
max 1.5 mA
Fuse protection in accordance with
national requirements
max. 16 A
Overvoltage category
II
Degree of soiling
2
Class of protection
I
CAUTION
Overloading of electrical cables
Electrical shock
▶
The total power of the connected electrical loads must not exceed 16 A (12
A in the case of UL versions), but can be distributed over the number of in-
stalled sockets.
▶
This limitation must also be ensured if 2 workstations are connected via
fuse protection (max. 16 A (12 A in the case of UL versions)).
▶
The same must be ensured if 3 workstations are connected via one power
supply cord.
Ambient conditions
Permissible installation sites
Interior rooms
Maximal installation height
2000 m above sea level
Permissible ambient temperature
range
+ 5
o
C to + 40
o
C
Permissible relative air humidity
max. 80 %
Dimensions
Height
945 or 870 mm
Adjustment range
30 mm
Width
1100 or 1200 mm
Single row depth
620 mm
Double row depth
1140 mm
Reduction of the maximum connec-
ted load per connected workplace
lamp or extra-low voltage supply
(with UL version)
1 A
13 / 36