CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the
sterilisation cycle!
The KaVo medical device has a maximum temperature resistance of up to
138 ℃ (280.4 °F).
Select a suitable procedure (depending on the available autoclave) from
the following sterilisation processes:
▪ Autoclave with three times pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
7 Reprocessing steps in accordance with ISO 17664
66 / 74