Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.3 Assembly of the KaVoLUX 540 LED U
4.3.4 Activate KaVoLUX 540 LED U in service mode
Note
The treatment light must be activated in service mode to activate the
KaVoLUX 540 LED U operating light on the treatment centres
ESTETICA E50/E50 Life/E70/E80.
Requirement
The following firmware versions are required for the operation of the
KaVoLUX 540 LED U:
Treatment
centre
SN
Unit
From Firmware
E50/E50 Life
-
Media-Gateway
V1.6.0
E70
200xxxxx
Unit
V2.3.0
Dentist
V2.1.3
Display Dentist
V2.3.0
E70
From 201xxxxx
Unit
V2.3.0
Dentist
V1.5.0
E80
100xxxxx
Unit
V2.3.0
Dentist
V2.1.3
Display Assist &
Dentist
V2.3.0
E80
From 101xxxxx
Unit
V2.3.0
Dentist
V1.5.0
KaVoLUX 540 LED
V1.2.1
▶
If necessary, carry out the firmware update.
See also:
2
TI of the respective treatment centre
ESTETICA E50/E50 Life, ESTETICA E70 from SN
20100000 and ESTETICA E80 from SN 10100000
Start service mode
▶
Press and hold down the "Function level" button.
▶
Press the "Remote Control" button.
Service mode appears on the display. If the service mode was active already,
this terminates the service mode.
▶
Press the "Remote Control" button until "U04" is shown on the display.