4 Operation
|
4.2 Filling spray water
4.2 Filling spray water
In order to avoid microbial contamination, the spray water reservoir must be reple‐
nished completely at least once a week.
If "SPRAY RESERVOIR EMPTY" is displayed on the touchscreen, this indicates
that the spray water reservoir is diminishing.
Fill the spray water as indicated below.
▶ Open flap ③ on the rear of the unit.
▶ Unscrew the spray water reservoir bottle ② in a clockwise direction.
▶ Fill spray water reservoir bottle.
Note
The spray water reservoir bottle may be filled only with distilled or demineralized
water (neutral pH). Acidic water may lead to corrosion in spray water supply system.
▶ Insert the aspiration tube ① again into the filled spray water reservoir bottle and
screw on the bottle by turning counterclockwise.
▶ Close door ③ so that the magnetic catch locks audibly.
Note
Alternatively, you can continue working immediately – with the remaining spray
water – by pressing the CONTINUE key on the touchscreen.