Note
If the pulse energy is changed when the laser is in ready mode, the KEY Laser 3+ /
III upgraded automatically switches to standby mode.
4.7.2
Select pulse frequency
Available: 2, 3, 4, 6, 10, 13, 15, 20, 25 and 30 Hz.
Note
If the pulse energy is changed when the laser is in ready mode, the KEY Laser 3+ /
III upgraded automatically switches to standby mode.
The selected setting is shown on the touchscreen (such as: 10 Hz).
The pulse frequency can be controlled with the arrow keys and
.
Alternately, the pulse frequency can also be set using the foot control by pressing
the button.
4.7.3
Display of average performance
Display of the set average performance in W
4.8
User test with handpiece 2060
Note
Make sure that the laser hose is inserted and screwed tight in the coupling, and
that handpiece 2060 is connected.
When the laser is in ready mode, the red pilot laser indicates the contact point of
the therapy laser beam. This pilot laser can be used to determine if the status of the
transmission system is okay.
See also: 4.6 Ready state, Page 37
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
4 Operation | 4.8 User test with handpiece 2060
40/76
Содержание KEY Laser 3+ 1343
Страница 1: ...Instructions for use KEY Laser 3 1343 KEY Laser III 1243 upgraded Always be on the safe side ...
Страница 79: ......
Страница 80: ...1 007 0175 kf 20100225 03 en ...