Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys-
28 / 44
The rinsing program proceeds in four steps. For rinsing programme 01, the
cycle starts in step 2, and for rinsing programme 02, the cycle starts in step 4.
Step
Description
4.0
Fresh water supply
Duration: 60 seconds
3.0
Tumbler rinsing
Duration: 20 seconds
2.0
Rinsing of the handpieces
The rinsing time can be set variably for each handpiece.
Standard setting:
▪ Rinsing programme 01 = 20 seconds per instrument
▪ Rinsing programme 02 = 2 minutes per instrument
1.0
Rinsing the triple-function and multifunction handpiece
Set rinsing time of rinsing programme 01 or 02
0.1
Waiting position for setting down the handpieces
Start rinsing programme
Note
The rinsing program can be started immediately or the next morning.
▶
Briefly press the "Intensive germ reduction" button to immediately start the
process.
ð
The selected rinse program starts.
or
▶
Turn off the device and do not restart it until the next morning.
ð
The cycle is started automatically. The rinsing time follows rinsing program
02.
Note
If the device is not used the entire day (i.e., the instruments are not moun-
ted), the rinsing program automatically starts the next morning when the
device is turned on.
The rinsing programme can be stopped at any time by pressing the "intensive
germ reduction" key.
If the device is turned off and then on again, the rinsing program also stops.
2.8.2 Intensive germ reduction
Note
Observe the instructions for use for the instruments.
Intensive germ reduction prevents the formation of microorganisms during ex-
tented periods when the equipment is out of use. During intensive germ reduc-
tion, the afferent water ducts are automatically filled with a higher concentra-
tion of KaVo OXYGENAL 6. The intensive germ reduction program lasts ca. 45
minutes. The increased hydrogen peroxide concentration acts in the system for
at least 30 minutes.