▶ Insert the test gauge (Mat. no. 04118731) between the hose and the
MULTIflex (LUX) / MULTIflex LED coupling and check the following
pressures:
ð
- Drive air: 2.5 to 3.0 bar (29 to 44 psi), recommended: 3.0 (44 bar psi)
ð
Return air: < 1 bar (15 psi)
ð
- Water: 1.0 to 2.0 bar (15 to 29 psi)
ð
- Spray air: 1.5 to 2.5 bar (22 to 36 psi)
4 Start up and shut down
34 / 88
Содержание INTRAflex LUX 2 program 2307 LN 0.535.6190
Страница 28: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture 3 Product description 28 88...
Страница 32: ...3 4 Start up and shut down 32 88...
Страница 86: ...9 Terms and conditions of warranty 86 88...
Страница 87: ......
Страница 88: ...1 005 3683 kb 20141204 07 en...