background image

DANGER

DANGER

indicates a hazardous situation that can directly cause death or serious

injury.

2.2   Safety instructions

WARNING

Hazards for the care provider and the patient.

In the case of damage, irregular running noise, excessive vibration, un-

typical warming or when the cutter or grinder cannot be held.

▶ Do not use further and notify Service.

 

Safety

9

Содержание INTRAcompact contra-angle 2068 LHC

Страница 1: ...Instructions for use INTRAcompact contra angle 2068 LHC REF 1 003 7701...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...ructions Warning symbol 7 2 1 2 Description of safety instructions Structure 7 2 1 3 Description of safety instructions Description of danger levels 8 2 2 Safety instructions 9 3 Product description 1...

Страница 4: ...grinders 25 5 4 Removing the milling tool or diamond grinder 29 6 Troubleshooting 31 6 1 Check for malfunctions 31 6 2 Troubleshooting 32 6 2 1 Troubleshooting Exchanging the O rings on the motor cou...

Страница 5: ...ual disinfection external 42 7 3 2 Disinfection Manual disinfection internal 43 7 3 3 Disinfection Machine disinfection external and internal 44 7 4 Drying 45 7 5 Care products and systems Servicing 4...

Страница 6: ...7 6 Packaging 52 7 7 Sterilisation 53 7 8 Storage 55 8 Tools 57 9 Terms and conditions of warranty 58 Table of contents 4...

Страница 7: ...aVo quality product By following the instructions below you will be able to work smoothly economically and safely Copyright by KaVo Dental GmbH Symbols Refer to Safety Warning symbol chapter Important...

Страница 8: ...equirements of the applicable EC directive Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable Target group This document is intended for dentists and their assistants The s...

Страница 9: ...symbol 2 1 2 Description of safety instructions Structure DANGER The introduction describes the type and source of the hazard This section describes the potential consequences of non observance The o...

Страница 10: ...ed here together with the three levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injuri...

Страница 11: ...h or serious injury 2 2 Safety instructions WARNING Hazards for the care provider and the patient In the case of damage irregular running noise excessive vibration un typical warming or when the cutte...

Страница 12: ...viced and stored in a dry location according to instructions before long periods of nonuse CAUTION Risk due to incorrectly stored instrument Injury and infection caused by chucked cutters or grinders...

Страница 13: ...Risk of burn injury from hot handpiece head or hot handpiece lid Burn injuries in the mouth may be caused if the instrument overheats Never touch soft tissue with the handpiece head or handpiece lid S...

Страница 14: ...dental treatment unit connected must have control equipment for changing the speed and direction of rotation A note is to be included in the documents accompanying the dental treatment unit referring...

Страница 15: ...ned by KaVo To ensure proper function the medical device must be set up according to the reprocessing methods described in the KaVo Instructions for Use and the care products and care systems describe...

Страница 16: ...Service may only be carried out by KaVo trained repair shops using original KaVo replacement parts Safety 14...

Страница 17: ...3 Product description INTRAcompact contra angle handpiece 2068 LHC Mat no 1 003 7701 Product description 15...

Страница 18: ...tment Any other type of use or alteration to the product is impermissible and can be hazardous The medical device is intended for the following uses Cavity preparation and caries excavation A medical...

Страница 19: ...e health and safety regulations the applicable accident prevention regulations these Instructions for use According to these regulations the user is required to only use equipment that is operating co...

Страница 20: ...tion 1 blue ring Speed transmission 1 1 With press button chuck Usable with contra angle cutters or grinders The contra angle handpiece can be mounted on all INTRAmatic motors and motors with a connec...

Страница 21: ...vice after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F T...

Страница 22: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 20...

Страница 23: ...er and patient Before first use and after each use sterilize the medical device 4 1 Cooling medium supply CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pul...

Страница 24: ...device during treatment A medical device that is not properly locked in place can become discon nected from the motor coupling and fall off Carefully pull on the medical device before each treatment...

Страница 25: ...y start the contra angle handpiece when the chuck is closed CAUTION Removing and attaching the contra angle handpiece while the drive motor is rotating Damage to the drive Never attach or remove the c...

Страница 26: ...he medical device to the LUX motor coupling until it locks into place and twist it until the catch audibly locks Pull on it to make sure that the medical device is securely attached to the coupling Op...

Страница 27: ...then pulling it along its axis 5 3 Insert the milling cutters or diamond grinders Note Only use carbide cutters or diamond burs that correspond to ISO 1797 1 type 1 are made of steel or tungsten carb...

Страница 28: ...tructions for use and use the cutter or grinder prop erly Only use cutters or grinders that do not deviate from the specified data CAUTION Injury from using worn cutters or grinders Cutters or grinder...

Страница 29: ...from defective chucking system The cutter or grinder could fall out and cause injury Pull on the cutter or grinder to check that the chucking system is okay and the cutter or grinder is securely held...

Страница 30: ...Insert the cutter or grinder into the segment of the head drive by twisting the tool slightly and push to the stop Check that the tool is seated by pulling on it Operation 28...

Страница 31: ...er Lacerations and damage to the chucking system Do not touch rotating cutter or grinder Never press the press button while the cutter or grinder is rotating Remove the cutter or grinder from the cont...

Страница 32: ...After the cutter or grinder has stopped rotating press the press button with your thumb and simultaneously pull out the drill bit or bur Operation 30...

Страница 33: ...g or damaged O rings Malfunctions and premature failure Make sure that all O rings are on the coupling and undamaged CAUTION Heating of the product Burns or product damage from overheating Do not use...

Страница 34: ...When the speed drops or is uneven Caring for the medical device An O ring is missing on the motor coupling Replace O ring 6 2 1 Troubleshooting Exchanging the O rings on the motor coupling CAUTION Ha...

Страница 35: ...r coupling may only be lubricated with a cotton ball wetted with KaVo spray Press the O ring between your fingers to form a loop Shove the O ring to the front and remove it Insert new O rings into the...

Страница 36: ...azard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check spray water channels and if necessary clean spray nozzles wit...

Страница 37: ...s Observe suitable personal protective measures Remove all residual cement composite or blood without delay Recondition the medical device as soon as possible after treatment Remove the cutter or grin...

Страница 38: ...ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 7 2 1 Cleaning Manual cleaning external Accessories required Tap water 30 C 5 C 86 F 10 F Brush e g medium hard too...

Страница 39: ...d with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher mediclean cleaning agent neodisher Z neu tralisation agent and...

Страница 40: ...dical device are dry and then lubricate immediately with care agents from the KaVo care system 7 2 3 Cleaning Manual cleaning internal Can only be done with KaVo CLEANspray or KaVo DRYspray Cover the...

Страница 41: ...al interior cleaning are only available in the following countries Germany Austria Switzerland Italy Spain Portugal France Luxem bourg Belgium Netherlands United Kingdom Denmark Sweden Fin land and No...

Страница 42: ...r mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the In...

Страница 43: ...ion CAUTION Malfunctioning from using a disinfectant bath or disinfectant containing chlorine Defects in the product Only disinfect in a thermodisinfector or manually Setup methods according to DIN EN...

Страница 44: ...n material compatibility The microbiological efficacy must be ensured by the disinfectant manufac turer Microcide AF from Sch lke Mayr liquid or cloths FD 322 from D rr CaviCide from Metrex Tools requ...

Страница 45: ...isinfection internal The efficacy of manual internal disinfection must be demonstrated by the manufacturer of the disinfection agent With KaVo products use only dis infection agents that have been rel...

Страница 46: ...d neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer...

Страница 47: ...compressed air until water drops are no longer visible Machine drying The drying procedure is normally part of the cleaning program of the ther modisinfector Follow the instructions for use of the the...

Страница 48: ...in the medical device Risk of injury from sharp or pointed cutters or grinders Remove cutter or grinder CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Pe...

Страница 49: ...7 5 1 Care products and systems Servicing Care with KaVo Spray KaVo recommends servicing the product after each time it is used i e after each automatic cleaning and before each sterilisation Remove c...

Страница 50: ...d servicing the chuck system once a week Remove the cutter or grinder place the spray nipple tip in the opening and spray Carry out the servicing according to the instructions in the section Care with...

Страница 51: ...sterilisation Place the product on the appropriate coupling of the KaVo SPRAYrotor and cover it with a CLEANpac bag Service the product See also Instructions for use KaVo SPRAYrotor 7 5 3 Care produc...

Страница 52: ...r each automatic cleaning and before each sterilisation Remove cutter or grinder Service the product Chuck care KaVo recommends cleaning and servicing the chuck system once a week See also Instruction...

Страница 53: ...r or grinder place the spray nipple tip in the opening and spray Subsequently treat with the care products and care systems specified See also Care with KaVo QUATTROcare Setup methods according to DIN...

Страница 54: ...tched The quality and use of the sterilised product packaging must satisfy ap plicable standards and be suitable for the sterilisation procedure Individually weld the medical device in the sterilized...

Страница 55: ...h EN 13060 ISO 17665 1 e g KaVo STERIclave B 2200 2200 P CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Before each sterilisation cycle service the medic...

Страница 56: ...sture Damage to product Immediately remove the product from the steam sterilizer after the sterilization cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Setup meth...

Страница 57: ...F 7 4 F Autoclave using the gravitation method at least 10 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F or alternatively at least 60 minutes at 121 C 1 C 4 C 250 F 1 6 F 7 4 F Use according to the manuf...

Страница 58: ...Comply with the expiry date of the sterilised items Setup methods according to DIN EN ISO 17664 56...

Страница 59: ...0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0921 Material summary Mat no KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spray...

Страница 60: ...r claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory legal reg...

Страница 61: ...m manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy...

Страница 62: ......

Страница 63: ......

Страница 64: ...1 003 7909 kb 20101028 03 en...

Отзывы: