Instructions for use

INTRAcompact contra-angle 2052 LHC - REF

1.003.7702

Содержание INTRAcompact contra-angle 2052 LHC

Страница 1: ...Instructions for use INTRAcompact contra angle 2052 LHC REF 1 003 7702...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...uctions 6 2 2 Safety instructions 8 3 Product description 13 3 1 Purpose Proper use 14 3 2 Technical Specifications 16 3 3 Transportation and storage conditions 17 4 First use 19 4 1 Checking the amou...

Страница 4: ...shooting 33 6 2 1 Troubleshooting Cleaning the spray nozzle 33 7 Preparation methods according to ISO 17664 35 7 1 Preparations at the site of use 35 7 2 Cleaning 36 7 2 1 Cleaning Manual cleaning ext...

Страница 5: ...4 Drying 45 7 5 Care products and systems Servicing 46 7 5 1 Care products and systems Servicing Care with KaVo Spray 47 7 5 2 Care products and systems Servicing Care with the KaVo SPRAYrotor 49 7 5...

Страница 6: ...aVo quality product By following the instructions below you will be able to work smoothly economically and safely Copyright by KaVo Dental GmbH Symbols Refer to Safety Warning symbol chapter Important...

Страница 7: ...equirements of the applicable EC directive Can be steam sterilised at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Thermodisinfectable Target group This document is intended for dentists and their assistants The s...

Страница 8: ...ions Warning symbol Structure DANGER The introduction describes the type and source of the hazard This section describes the potential consequences of non observance The optional step includes necessa...

Страница 9: ...the three levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injuries WARNING WARNING in...

Страница 10: ...h or serious injury 2 2 Safety instructions WARNING Hazards for the care provider and the patient In the case of damage irregular running noise excessive vibration un typical warming or when the cutte...

Страница 11: ...dental treatment unit connected must have control equipment for changing the speed and direction of rotation A note is to be included in the documents accompanying the dental treatment unit referring...

Страница 12: ...viced and stored in a dry location according to instructions before long periods of nonuse CAUTION Risk due to incorrectly stored instrument Injury and infection caused by chucked cutters or grinders...

Страница 13: ...ned by KaVo To ensure proper function the medical device must be set up according to the reprocessing methods described in the KaVo Instructions for Use and the care products and care systems describe...

Страница 14: ...Service may only be carried out by KaVo trained repair shops using original KaVo replacement parts Safety 12...

Страница 15: ...3 Product description INTRAcompact contra angle handpiece 2052 LHC Mat no 1 003 7702 Product description 13...

Страница 16: ...tment Any other type of use or alteration to the product is impermissible and can be hazardous The medical device is intended for the following uses Cavity preparation and caries excavation A medical...

Страница 17: ...e health and safety regulations the applicable accident prevention regulations these Instructions for use According to these regulations the user is required to only use equipment that is operating co...

Страница 18: ...tification 1 blue ring Speed transmission 1 1 With press button chuck Miniature drill bits can be used The contra angle handpiece can be mounted on all INTRAmatic motors and motors with a connection i...

Страница 19: ...vice after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F T...

Страница 20: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 18...

Страница 21: ...first use and after each use sterilise the medical device CAUTION Damage from soiled and moist cooling air Contaminated and moist cooling air can cause malfunctions and lead to premature bearing wear...

Страница 22: ...cking the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the dental pulp Adjust the water amount for the spray cooling to a minimum of 50 cm3 min First...

Страница 23: ...ent spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 09...

Страница 24: ...device during treatment A medical device that is not properly locked in place can become discon nected from the motor coupling and fall off Carefully pull on the medical device before each treatment...

Страница 25: ...attach or remove the contra angle handpiece while the drive motor is rotating CAUTION Damage If the footswitch is pressed while you are connecting and removing the medical device you can damage the me...

Страница 26: ...the motor coupling with KAVOspray Attach the medical device to the motor coupling and twist it until the guide stud audibly locks Pull on the medical device to make sure that it is securely affixed t...

Страница 27: ...rom the motor coupling by twisting it slightly and then pulling it along its axis 5 3 Inserting the miniature bur Note Use only miniature burs whose shafts comply with ISO 1797 1 type 1 Shaft diameter...

Страница 28: ...r damage to the medical device Only use miniature burs that do not deviate from the specified data CAUTION Injury from using worn miniature burs Miniature burs can fall out during treatment and harm t...

Страница 29: ...m defective chucking system The miniature bur can fall out and cause injury Pull on the miniature bur to check if the chucking system is working properly and securely holds the miniature bur Wear glov...

Страница 30: ...Insert the miniature bur into the segment of the head drive by twisting the it slightly and then push it to the stop Check that the tool is seated securely by pulling on it Operation 28...

Страница 31: ...Lacerations and damage to the chucking system Do not touch rotating miniature bur Never actuate the push button while miniature bur is rotating Remove the miniature bur from the contra angle handpiece...

Страница 32: ...After the miniature bur has stopped rotating firmly press the push but ton with your thumb and pull out the miniature bur Operation 30...

Страница 33: ...g or damaged O rings Malfunctions and premature failure Make sure that all O rings are on the coupling and undamaged CAUTION Heating of the product Burns or product damage from overheating Do not use...

Страница 34: ...heck the amount of cooling air The medical device is too hot while working Caring for the medical device When the speed drops or is uneven Caring for the medical device An O ring is missing on the mot...

Страница 35: ...spray nozzle CAUTION Hazard from insufficient spray water Insufficient spray water can cause the medical device to overheat and damage the tooth If the amount of spray water is insufficient unblock th...

Страница 36: ...Use the nozzle needle Mat no 0 410 0921 to free the water passage in the spray nozzles Troubleshooting 34...

Страница 37: ...inated medical devices Observe suitable personal protective measures Remove all residual cement composite or blood without delay The medical device must be dry when transporting it to the site of re p...

Страница 38: ...ions from cleaning in an ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 7 2 1 Cleaning Manual cleaning external Accessories required Tap water 30 C 5 C 86 F 10 F...

Страница 39: ...d with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher mediclean cleaning agent neodisher Z neu tralisation agent and...

Страница 40: ...dical device are dry and then lubricate immediately with care agents from the KaVo care system 7 2 3 Cleaning Manual cleaning internal Can only be done with KaVo CLEANspray or KaVo DRYspray Cover the...

Страница 41: ...ual interior cleaning are only available in the following countries Germany Austria Switzerland Italy Spain Portugal France Luxem bourg Belgium Netherlands United Kingdom Denmark Sweden Fin land and N...

Страница 42: ...er mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the I...

Страница 43: ...ion CAUTION Malfunctioning from using a disinfectant bath or disinfectant containing chlorine Defects in the product Only disinfect in a thermodisinfector or manually Preparation methods according to...

Страница 44: ...on material compatibility The microbiological efficacy must be ensured by the disinfectant manufac turer Microcide AF from Sch lke Mayr liquid or cloths FD 322 from D rr CaviCide from Metrex Tools req...

Страница 45: ...isinfection internal The efficacy of manual internal disinfection must be demonstrated by the manufacturer of the disinfection agent With KaVo products use only dis infection agents that have been rel...

Страница 46: ...G 7881 validation was performed with the VARIO TD programme neodisher mediclean cleaning agent neodisher Z neu tralisation agent and neodisher mielclear rinsing agent and extends only to the compatibi...

Страница 47: ...with care agents from the KaVo care system 7 4 Drying Manual drying Blow off the outside and inside with compressed air until water drops are no longer visible Machine drying The drying procedure is n...

Страница 48: ...medical device Risk of injury from sharp or pointed cutters or grinders Remove cutter or grinder CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Perform r...

Страница 49: ...7 5 1 Care products and systems Servicing Care with KaVo Spray KaVo recommends servicing the product after each time it is used i e after each automatic cleaning and before each sterilisation Remove c...

Страница 50: ...g the chuck system once a week Remove the cutter or grinder place the spray nipple tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaVo Spra...

Страница 51: ...each sterilisation Place the product on the appropriate coupling of the KaVo SPRAYrotor and cover it with a CLEANpac bag Service the product See also Instructions for use KaVo SPRAYrotor 7 5 3 Care p...

Страница 52: ...r each automatic cleaning and before each sterilisation Remove cutter or grinder Service the product Chuck care KaVo recommends cleaning and servicing the chuck system once a week See also Instruction...

Страница 53: ...er or grinder place the spray nipple tip in the opening and spray Subsequently treat with the care products and care systems specified See also Care with KaVo QUATTROcare Preparation methods according...

Страница 54: ...tched The quality and use of the sterilised product packaging must satisfy ap plicable standards and be suitable for the sterilisation procedure Individually weld the medical device in the sterilised...

Страница 55: ...th EN 13060 ISO 17665 1 e g KaVo STERIclave B 2200 2200 P CAUTION Premature wear and malfunctions from improper servicing and care Reduced product life Before each sterilisation cycle service the medi...

Страница 56: ...sture Damage to product Immediately remove the product from the steam steriliser after the sterilisation cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Preparatio...

Страница 57: ...nitial vacuum at least 3 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F Autoclave using the gravitation method at least 10 minutes at 134 C 1 C 4 C 273 F 1 6 F 7 4 F or alternatively at least 60 minutes a...

Страница 58: ...processed products should be stored protected from dust with min imum exposure to germs in a dry dark and cool space Comply with the expiry date of the sterilised items Preparation methods according t...

Страница 59: ...9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0921 Material summary Mat no KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spray 2112...

Страница 60: ...r claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory legal reg...

Страница 61: ...rom manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a cop...

Страница 62: ......

Страница 63: ......

Страница 64: ...1 003 7924 kb 20101006 02 en...