Technician's instructions KaVo In eXam 3510
2 Complete exchange of the x-ray head
|
2.1 Remove the x-ray head
2 Complete exchange of the x-ray head
2.1 Remove the x-ray head
Note
Disconnect the In eXam from the mains, and make sure that it cannot be turned
on.
▶ Remove the tube ①.
▶ Remove the lock screws in the x-ray head cover, and remove the cover ②
1
2
▶ Unplug the laser diode ③
▶ Unplug the x-ray head from the arm cable ④
▶ Place the scissor arm in a vertical position, and secure it with belts.
Injury from unsecured scissor arm.
18/32