▶ Weld each medical device individually in a sterilised item package!
7.7
Sterilisation
Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Before each sterilisation cycle, service the medical device with KaVo care
products.
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisa‐
tion cycle!
The medical device has a maximum temperature resistance up to 138℃
(280.4°F).
KaVo recommends for example
▪ STERIclave B 2200/ 2200P by KaVo
▪ Citomat/ K-series by Getinge
Autoclave with a pre-vacuum for least four minutes at 134°C ± 1°C
(273°F ± 1.8°F)
Drying time: 20 min.
Autoclave with the gravitation method for at least 10 minutes at 134℃ ±1°C ± 1°C
(273°F ± 1.8°F)
Drying time: 30 min.
Autoclave with the gravitation method for at least 60 minutes at 121° ±1°C
(250°F ± 1.8°F)
Drying time: 30 min.
Follow the manufacturer's instructions for use.
7.8
Storage
Prepared products must be stored, protected from germs (as far as possible) and
dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.
Instructions for use GENTLEsilence 8000BN - 1.000.5700
7 Preparation methods according to ISO 17664 | 7.7 Sterilisation
18/19