6.2.3 Cleaning: Manual cleaning - internal
Validated manual internal cleaning (removal of residual protein) can only
be done with KaVo CLEANspray and KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three
times for 2 seconds each time. Remove the medical device from the
spray attachment and let the cleaner work for one minute.
▶
Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also: KaVo CLEANspray / KaVo DRYspray Instructions for Use
Preparation methods according to ISO 17664
35
Содержание GENTLEpower LUX handpiece 10 LP
Страница 1: ...Instructions for use GENTLEpower LUX handpiece 10 LP 1 001 5660...
Страница 6: ...7 Tools and consumables 61 8 Warranty terms and conditions 63 Table of contents 4...
Страница 15: ...3 Product description GENTLEpower LUX Handpiece 10 LP Mat no 1 001 5660 Product description 13...
Страница 20: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 18...
Страница 23: ...3 Start up and shut down 21...
Страница 62: ...Note Comply with the expiry date of the sterilised items Preparation methods according to ISO 17664 60...
Страница 67: ......
Страница 68: ...1 001 9319 kb 20130802 06 en...