6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions

16 / 28

6 Checking for malfunctions and troubleshooting

6.1 Checking for malfunctions

 CAUTION

Overheating of the device.

Burns or product damage from overheating.

4

If the device overheats, stop working and have the service personnel repair
the device.

CAUTION

Missing or damaged O-rings.

Malfunction and premature failure.

4

Make sure that all O-rings are on the coupling and are undamaged.

4

If the medical device overheats while idling, check the cooling air flow.

4

If the medical device overheats when exposed to load, the medical device
needs to be serviced.

4

When the speed drops or is uneven, the medical device needs to be ser-
viced.

4

If there is no O-ring on the motor coupling, replace the O-ring.

Also refer to:

Instructions for use of motor

6.2 Troubleshooting

 WARNING

Repair WITHOUT using KaVo original spare parts.

Parts could come loose and cause injury. Aspiration, swallowing of parts and
danger of suffocation.

4

Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.

NOTE

If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.

The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by MPG (in English medical device law) §4, section 1 No. 1 and there-
fore requires its own conformity check.

6.2.1 Replacing the O-rings on the motor coupling

CAUTION

Improper care of the O-rings.

Malfunction or complete failure.

4

Do not use Vaseline or other grease or oil.

4

Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the
coupling.

4

Press the O-ring between your fingers to form a loop.

4

Push the O-ring to the front, and remove it.

Содержание EXPERTmatic E31 C

Страница 1: ...Instructions for use EXPERTmatic E31 C 1 008 4998...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Страница 3: ...g for malfunctions and troubleshooting 16 6 1 Checking for malfunctions 16 6 2 Troubleshooting 16 6 2 1 Replacing the O rings on the motor coupling 16 7 Reprocessing steps in accordance with ISO 17664...

Страница 4: ...Vo Technical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use ar...

Страница 5: ...storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original langu...

Страница 6: ...ns for use EXPERTmatic E31 C 1 008 4998 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to materi...

Страница 7: ...Reprocess the product and accessories appropriately before disposal 4 Use gloves or a finger guard when you test insert and remove the dental bur 2 2 Improper use Because the operation with an electri...

Страница 8: ...sories that are equipped with standardised interfaces 4 Do not make any modifications to the device unless these have been ap proved by the manufacturer of the product 4 Use original KaVo spare parts...

Страница 9: ...r third parties This may res ult in aspiration swallowing of parts and possibly even a risk of suffocation 4 Only use spare parts that comply with the specification for repair original manufacturer sp...

Страница 10: ...r the described application by a properly trained user You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructi...

Страница 11: ...speed via the motor control 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm u...

Страница 12: ...uct and accessories before disposal Also refer to 7 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 18 Current packaging law Dispose of and recycle the packaging appropriately in ac...

Страница 13: ...securely locked onto the coupling CAUTION Removing and attaching the medical device while the drive motor is ro tating Damage to the catch pin 4 Never attach or remove the medical device while the dr...

Страница 14: ...ird parties could be infected by contaminated medical devices 4 Use gloves or a finger guard when you test insert and remove the dental bur CAUTION Dental bur with damaged worn or deformed shafts Mate...

Страница 15: ...c E31 C 1 008 4998 5 Operation 5 4 Removing the dental bur 15 28 4 Unscrew it or take it off the insert after it has stopped rotating NOTE For the re use of the inserts please refer to the instruction...

Страница 16: ...HOUT using KaVo original spare parts Parts could come loose and cause injury Aspiration swallowing of parts and danger of suffocation 4 Only use spare parts that comply with the specification for repa...

Страница 17: ...for use EXPERTmatic E31 C 1 008 4998 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting 17 28 4 Spray a lint free cloth with KaVo Spray apply it to the new O rings and in sert them in...

Страница 18: ...only KaVo recommends the following products based on the compatibility of the ma terials The microbiological efficacy must be ensured by the disinfectant manu facturer and proven by an expert opinion...

Страница 19: ...internal and external disinfection Manual cleaning Accessories required Tap water drinking water Brush e g medium hard toothbrush 4 Brush off under flowing tap water at least 10 seconds Cleaning with...

Страница 20: ...only CAUTION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only 7 3 1 pre cleaning Accessories required T...

Страница 21: ...ng with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation but no later than after 30 minute...

Страница 22: ...F Sterilisation parameters Select a suitable process from the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1...

Страница 23: ...ry Mat no Holder 1 004 0596 INTRA Instrument stand 3 005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spr...

Страница 24: ...liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connectio...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ...1 008 5577 bd 20240417 03 en...

Отзывы: