KaVo EWL 5831 Скачать руководство пользователя страница 8

6

EWL 5831

A 6 Installation

A 6.1 Location

As the unit uses primarily water, installation is recommended close to a water tap,  preferably at a sink, or wall-mounted on a bracket above
the sink.   

A 6.2 Water connection

Connect the reinforced pressure hose between the faucet and the unit connector,  located at the rear under the motor.  Firmly secure the hose
using the pre-installed clips.

The faucet should be closed when the unit is not in operation, in order to prevent the hose standing permanently under pressure.

• Re-tighten the hose clips about 2 days after installation.
• The water supply must be shut off overnight and at weekends.

A 6.3 Drainage hose

The drainage hose must be so placed that, under no circumstances, a back-up of water occurs at the unit (spraying water). Lay the hose in a
smooth curve (not kinked), with a constant drop of at least 1 cm per metre (1%).

The end of the drainage hose must remain above the water level of the plaster trap, and never be submerged in water.

As standard, the hose connector is set diagonally to the right  However, this can be converted to a straight connector, which is
recommended under difficult drainage conditions.
In this case, simply exchange the sealing cap for the connector.

Содержание EWL 5831

Страница 1: ...structions Model Trimmer EWL 5831 1 001 3277 RB 05 02 GB 15 07 KaVo Elektrotechnisches Werk GmbH Wangener Stra e 78 D 88299 Leutkirch Tel 0 75 61 86 0 Fax 0 75 61 86 22 2 1 000 8422 DE 1 000 9674 GB 1...

Страница 2: ......

Страница 3: ...enance 4 A 2 2 Cleaning 4 A 3 Applications 4 A 4 Preparing for installation 4 A 4 1 Unpacking Transport 4 A 4 2 Included accessories 4 A 4 3 Setting up and location 4 A 5 Handling 5 A 5 1 Technical da...

Страница 4: ...the motor cannot be restarted When working with the trimming disc safety glasses must be worn Electric motors operate with dangerous electrically charged rotating parts By incorrect operation improper...

Страница 5: ...o EWL are not extended through the above information Modifications by others and use of unauthorized parts invalidates the Product Liability Use only original KaVo EWL spare parts After initial instal...

Страница 6: ...cation A 4 Preparing for installation A 4 1 Unpacking Transport Open the shipping carton remove the packing materials and carefully take out the unit and accessories Check the included accessories Whe...

Страница 7: ...ht of 1 60 metres In a time related average of 50 free running 30 medium load and 20 full load these amounted to 69 db A The noise recordings were carried out in accordance with DIN 45635 Part 1 Wait...

Страница 8: ...sure Re tighten the hose clips about 2 days after installation The water supply must be shut off overnight and at weekends A 6 3 Drainage hose The drainage hose must be so placed that under no circums...

Страница 9: ...Please note the following When working with the diagonal support ensure that it is correctly located according to the sketch below When fitting the diagonal support make sure that positions 1 and 2 ar...

Страница 10: ...rp Make sure the hose lies as straight as possible The hose is fitted with a pipe elbow Curve of the elbow must be at least 40 mm A 8 3 Reconditioning the trimming disc After a period of time if the t...

Страница 11: ...ements free of charge KaVo EWL is not liable for defects and their consequences caused by or which could be caused by normal wear misuse improper cleaning or maintenance failure to follow the operatin...

Страница 12: ...mal dia 260 grain size 24 1 000 8873 Replacement diamond disc 1 001 2202 Tool for changing discs 1 001 2203 Sealing grease SPARE PARTS EWL 5831 10 EWL 5831 Spare parts 1 2 1 001 3426 1 001 3427 1 001...

Страница 13: ...GB OPERATING INSTRUCTIONS EWL 5831 11...