▶ Fill the intensive cleaning bottle.
▶ Press key while flushing.
▶ Flush each handpiece for 20 seconds (without attachment handpieces or cannulas), including the
three-function syringe.
▶ Actuate the tumbler for 15 seconds.
▶ Then allow to act for at least 30 minutes, and preferably over the weekend.
▶ After the time of exposure, e.g. after the weekend, actuate the tumbler for 15 seconds and rinse
each handpiece (without attachment handpieces or cannulas) for 20 seconds.
See also: Servicing Instructions ESTETICA E30
8.4
Conditioning the suction system
Note
Clean the suction hoses after each treatment and disinfect them with DEKASEPTOL
Gel daily.
Application of DEKASEPTOL Gel
Note
Regarding the air jet suction system (Venturi), only one aliquot may be drawn into
each hose.
▶ Aspirate one tumbler of cold water with each suction hose.
▶ With the suction handpiece attached or by hand, press down the DEKASEPTOL dosing facility once
each for the saliva ejector hose and the spray mist ejector.
▶ Aspirate the DEKASEPTOL Gel from the dosing facility.
Note
In suction processes with air jet suction systems (Venturi), the two following steps
must not be carried out as they could lead to blockage of the thin hoses.
Hygiene