Instructions for use DSE Expert
6 Service and Care | 6.1 Servicing
60 / 74
① Wastewater tank
② Pressure tank with holding magnets
▶ Detach the empty pressure tank ② only by a few mm from the holding magnets
until the pressure in the pressure tank decreases. Detach the right magnet first by
holding the pressure tank by its handle and rotating it a little toward the left. Then
remove the pressure tank. The pressure tank is affixed by 2 holding magnets.
▶ Unscrew the lid, add Oxygenal 6 and water. Screw the lid on again.
▶ Insert the pressure tank ② along the right inside of the aluminium profile until the
magnets can be heard to click.
6.1.3 Permanent germ reduction and intensive germ reduction in the
presence of the integrated pressure tank (if available)
CAUTION
Wrong mixing ratio.
Too much or too little Oxygenal 6.
▶ The bottle must always be empty before adding Oxygenal 6.
The DSE Expert can be operated optionally with a permanent germ reduction and an
intensive germ reduction. This is made possible by dosing KaVo OXYGENAL 6 into
the pressure tank.
Permanent germ reduction
Permanent germ reduction with KaVo OXYGENAL 6 reduces the formation of algae,
germs and biofilms and provides germ-free treatment water in accordance with the
drinking water regulations for safe health of the students.
Oxygenal concentrations
Feature
Percentage fraction
Amount in ml per 1 L water
Permanent germ reduction max. 0.02 %
3 ml