Instructions for use DEXIS CariVu 3-in-1 by KaVo
7 Processing steps in accordance with ISO 17664 | 7.1 Overview of the processing options
7 Processing steps in accordance with ISO 17664
7.1 Overview of the processing options
Components Single use
or Multi-
person use
Manual
Cleaning
Manual Dis-
infection
Automated
Cleaning
Automated
Disinfection
Sterilized
DEXIS CariVu
3-in-1 by
KaVo
Multi-Person
use
Yes
Yes
No
No
No
Tip Vision Full
HD
Multi-Person
use
Yes
No
Yes
Yes
Yes
Fluorescence
Cover Vision
Full HD
Multi-Person
use
Yes
No
Yes
Yes
Yes
Cradle Vision
Full HD
Multi-Person
use
Yes
Yes
No
No
No
Wall Mount
Vision Full HD
Multi-Person
use
Yes
Yes
No
No
No
Sheath Vision
Full HD
Single Use
No
No
No
No
No
7.2 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection risk.
▶
Take suitable personal protective measures.
▶
Pull off the USB plug of the medical device.
▶
Process the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to processing.
▶
To minimize the risk of infection during processing, always wear protective
gloves.
▶
Remove gross soiling immediately.
▶
Do not immerse in solutions or the like.
▶
Repeat cleaning if the device is not visually clean at the end of the cleaning
step. If device cannot be visually cleaned then safely dispose of the device.
37 / 50