7.5.5
Servicing with KaVo QUATTROcare 2104 / 2104A
Note
Can be used with the COMPACTtorque 636C only. The COMPACTtorque 636
cannot be fitted in the QUATTROcare!
Cleaning and servicing unit with expansion pressure for effective cleaning and
care.
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶ Remove the cutter or grinder.
▶ Servicing the product.
Servicing the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
See also: Instructions for use KaVo QUATTROcare 2104 / 2104A
▶ Remove the cutter or grinder, place the spray nipple tip in the opening and
spray.
▶ Subsequently treat with the care products and care systems specified.
See also: Servicing with KaVo QUATTROcare 2104 / 2104A
7.5.6
Servicing with KaVo QUATTROcare Plus
Note
Can be used with the COMPACTtorque 636C only. The COMPACTtorque 636
cannot be fitted in the QUATTROcare!
KaVo recommends servicing the product after each time it is used, i.e. after each
automatic cleaning and before each sterilisation.
▶ Remove the cutter or grinder.
▶ Servicing the product in QUATTROcare PLUS.
See also: Instructions for Use KaVo QUATTROcare PLUS 2124 A
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830
7 Reprocessing methods according to ISO 17664 | 7.5 Care products and systems – Servicing
18/22