Instructions for use SMARTmatic S20 - 1.011.6750 SMARTmatic S20 K - 1.011.6751 SMARTmatic S20 S - 1.011.6752
4 Startup and shut down | 4.1 Checking the amount of water for the S20 S
4 Startup and shut down
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶ Prior to initial startup and after each use, process the product and accessories.
WARNING
Dispose of the product in the appropriate manner.
Infection hazard.
▶ Process the product and accessories before disposal.
NOTICE
Damage from contaminated and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
▶ Make sure that the supplied cooling air is dry, clean and free of contamination in
accordance with ISO 7494-2.
4.1 Checking the amount of water for the S20 S
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Thermal damage to the dental pulp.
▶ Adjust the water amount for the spray cooling to a minimum of 50 ml/min!
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage the
tooth.
▶ Check the spray water channels and clean the spray tube with the nozzle needle
(Mat. no. 0.410.0931) according to need.