Instructions for use SURGmatic S201 XL - 1.010.2332 SURGmatic S201 XC - 1.010.2333
6 Troubleshooting | 6.1 Check for malfunctions
6 Troubleshooting
6.1 Check for malfunctions
CAUTION
Heating of the product.
Burns or product damage from overheating.
▶
Do not use the product if it is irregularly heated.
▶
The medical device is too hot while working:
Service the medical device.
▶
When the speed drops or is uneven:
Service the medical device.
▶
An O-ring is missing on the motor coupling:
Replace O-ring.
See also:
2
Instructions for use of motor
6.2 Troubleshooting
6.2.1 Cleaning the spray clip and the spray tube
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check spray water channels and if necessary clean spray tubes with the
nozzle needle (
Mat. no. 0.410.0931
).
15 / 26