Troubleshooting
36
Exchanging the O-rings on the coupling on the supply hose
CAUTION
Hazard from improper care of the O-rings.
Improper care of the O-ring can cause the medical device to partially or
completely malfunction.
▶
Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wet
with KAVOspray.
▶
Press the O-ring between your fingers to form a loop.
▶
Shove the O-ring to the front, and remove it.
▶
Insert new O-rings into the grooves.
Содержание 1.006.5500
Страница 1: ...Instructions for use GENTLEsilence LUX gold 8000B REF 1 006 5500...
Страница 6: ...Contents 4...
Страница 19: ...Product description 17 Product description GENTLEsilence LUX gold 8000B Mat no 1 006 5500...
Страница 40: ...Troubleshooting 38 Clean the water passage in the spray nozzles by using the nozzle needle Mat no 0 410 0921...
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ...1 007 1382 kb 20090227 01 en...