Autoclave with a triple pre-vacuum for at least 4 minutes at 134 °C ± 1 °C (273 °F ± 1.8 °F)
Autoclave with gravitation process for at least 10 minutes at 134 °C ± 1 °C (273 °F ± 1.8 °F)
Autoclave with gravitation process for at least 60 minutes at 121°C ± 1°C (250°F ± 1.8°F)
Comply with the manufacturer's instructions for use with regard to the application range.
6.8
Storage
Prepared products must be stored, protected from germs (as far as possible) and dust, in a dry,
dark, cool room.
Reprocessing methods according to ISO 17664
69
Содержание 1.003.7714
Страница 1: ...Instructions for use INTRAcompact contra angle 2061 CHC REF 1 003 7714 ...
Страница 7: ...6 8 Storage 69 7 Tools and consumables 71 Content 5 ...
Страница 14: ...2 Safety 2 1 Description of safety instructions Warning symbol Safety 12 ...
Страница 26: ...3 Product description INTRAcompact contra angle handpiece 2061 CHC Mat no 1 003 7714 Product description 24 ...
Страница 34: ... Place the medical device on the LUX motor coupling and lock it into place Operation 32 ...
Страница 36: ...5 3 Inserting the file Note Manufacturer of files Intensiv Operation 34 ...
Страница 38: ...CAUTION Danger of injury from files Infections or cuts Wear gloves or finger stalls Operation 36 ...
Страница 40: ... Turn the knurled wheel to fix the file 36 individual locking positions can be selected Operation 38 ...
Страница 42: ... When the file has stopped moving push file out of the head with ejector and remove it Operation 40 ...
Страница 49: ... Brush off under flowing tap water Reprocessing methods according to ISO 17664 47 ...
Страница 72: ...Note Comply with the expiry date of the sterilised items Reprocessing methods according to ISO 17664 70 ...
Страница 75: ......
Страница 76: ...1 003 8083 kb 20101118 02 en ...