64
1.
Disconnect the instrument from the wall outlet.
2.
Remove the screws marked below (or symmetrically, if you are remove the
other side’s panel).
3.
Slide the panel upwards to remove it.
Bottom panels
removal
Under the zone where the platinumware holders swing in and out of the furnace,
this fluxer has two panels that act as a protective shield for the electronics
underneath. It is necessary to remove those panels to access the solution
magnetic stirrer, mold cooling blowers and other electronic components.
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
Front bottom panel
The “front bottom panel” is the one with cutouts for the mold cooling blowers.
1.
Remove the top/back panel (optional).
2.
Open the safety shield.
Содержание X-FLUXER