61
2.
While a heating element can be replaced on an X-300 that is still hot (using
appropriate protective gloves)
Katanax strongly recommends letting the
unit cool down completely overnight.
3.
Remove the instrument’s top/back panel. (See page 62,
removal for details.)
4.
Remove the instrument’s LED/safety guard plate (2 screws).
5.
Loosen the two screws of the element stabilizer clamp.
6.
Proceed to remove one element at a time.
7.
Lower (unscrew) the stopper screw corresponding to the element position you
want to remove. Do not completely remove the screw; just loosen it until the
threads are lowered flush.
8.
Loosen the four (4) screws of the ceramic terminal connector.
LED/safety guard
Stopper screws
Clamp screws
Stabilizer clamp
Clamp screws