84
3.
Remove the blower box screws: three (3) on the front, three (3) on the top
towards the rear, and two (2) on the sides where the linear motion system
runs (see picture below).
Base
compartment
access
To clean or maintain the linear motion system that pulls the platinumware holder
in and out of the furnace, one must open the base of the instrument.
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
5.
Disconnect the instrument from the wall outlet.
6.
Flip the furnace backwards (see instructions at page 82).
7.
Remove the screws from the base top panel (see picture below).