82
WARNING
: Heating elements are powered at high voltage. Always disconnect
the instrument from the wall outlet before removing door. When no door is
present, safety interlocks are bypassed.
1.
Disconnect the instrument from the wall outlet.
2.
Lift the safety shield securely up/backwards.
3.
Ask for the assistance of a helper.
4.
While the helper is maintaining the door lever, remove its three (3) front
screws.
5.
With the helper now maintaining both levers, remove other side’s screws.
6.
Have your helper slowly accompany the levers’ upwards movement until the
levers stop.
Furnace
backwards
flipping
To access components located in the base, it is necessary to flip the furnace to the
back:
1.
Disconnect the instrument from the wall outlet.
2.
Lay the instrument on a surface large enough to accommodate the back face
of the furnace when you will flip it back.
3.
Manually open the door, and pull the platinumware holders out of the furnace.