Shutting the system down
After every treatment, the device must be cleaned and disinfected depending on application and
in accordance with the specifications in 12 Cleaning and maintenance. The device must then be
moved into the transport position described in 10.1 Transport position / idle position. This is the
optimal idle position when the device is not in use. If the device is not used for longer than 24
hours, putting on the lens covers (supplied) and pulling the protective cover (supplied) over the
device are recommended. Similarly, unused accessories should be removed and placed into the
storage packaging provided. This prevents damage to the lenses and dirtying of the device.
12
Cleaning and maintenance
12.1
Fuse replacement
Carry out the following steps as in Figure 21:
Unplug the mains connector. Insert a screwdriver into the slit of the fuse holder (1) and turn it
anticlockwise by 90°. A spring presses the cap up.
Remove the cap and replace the fuse attached in the cap
Insert the cap with a new fuse in, press it down and lock it into place by turning the
screwdriver clockwise by 90°
Figure 21
CAUTION
Only replace defective fuses with those having the same ratings
(nominal voltage, nominal current and switch-off characteristics).