TABmaster Model 6715 

 

Instruction Manual

 

 

 

 
 
 

 
 

 
 

 

 

 

 
 

 

 

 

 
 

 

 

 

Be sure to read this manual thoroughly before using the instrument , 

fully understand and pay attention to each cautions mentioned. 

Please well-keeping this manual for long time service reference

 
 

 
 

01003 

14.10 

Содержание TABmaster 6715

Страница 1: ...Instruction Manual Be sure to read this manual thoroughly before using the instrument fully understand and pay attention to each cautions mentioned Please well keeping this manual for long time service reference 01003 14 10 ...

Страница 2: ... 3 3 Pitot tube 11 3 4 Micromanometer 11 3 5 Shoulder Strap 11 4 Operating Instruction 12 4 1 Power supply 12 4 2 Start On Power Off 12 4 3 Keypad operation 13 5 Function test 14 5 1 User Interface 14 5 2 Single Mode 15 5 3 Average Mode 15 5 4 Back Pressure Mode 15 6 Menu setting 16 6 1 General setting 17 6 1 1 Date 17 6 1 2 Time 17 6 1 3 Auto Off 17 6 1 4 Backlighting 18 6 1 5 Communication 18 6 ...

Страница 3: ... 2 ID application 20 6 2 3 Test Mode 20 6 2 5 Data record 21 6 2 6 Standard Actual test 21 6 2 7 Compensation factor 22 7 Error and Troubleshooting 23 8 Warranty and Service 24 8 1 Product Warranty 24 8 2 After service 24 ...

Страница 4: ...ual and software inside 1 Quick Guide 1 6715 Optional Parts Item Specifications Spare hood 2x2ft Spare hood 2x4ft Spare hood 1x4ft Spare hood 3x2ft Spare hood 3x3ft Spare hood 500x500mm Poles Capture Hood Stand Up to 3 4m with capture hood Communication cable Printer AC adapter DC5V International plug adapter Carrying case Velocity Grid Pitot tube 2 3 x 200mm Pitot tube 4 x 300mm Pitot tube 8 x 50...

Страница 5: ...e 1 Calibration certificate 1 CD Rom Manual and software inside 1 Quick Guide 1 6700 Optional Parts Item Specifications Velocity Grid Pitot tube 2 3 x 200mm Pitot tube 4 x 300mm Pitot tube 8 x 500mm Pressure tubing 2 250cm Shoulder Strap Printer AC adapter DC5V International plug adapter Communication cable ...

Страница 6: ...ndicates a mandatory action A specific action is given near the symbol WARNING Heated forbidden Never bring the fabric hood to a flammable gas atmosphere otherwise the heat may cause a fire or explosion Disassemble prohibition Do not disassemblyor refit the instrument otherwise may cause the electric shock or a fire Using properly Using properlyunder the instruction manual otherwise may cause sens...

Страница 7: ... damage Forbidden Do not wipe the instrument with a volatile solvent the body may deform or deteriorate Use a soft dry cloth to remove stains If stains persist soak the cloth in a neutral detergent and wipe the instrument with the soft cloth Never use volatile solvents such as thinner or benzene Forbidden Do not use or leave the instrument in a high temperature high humidity or dusty environment D...

Страница 8: ...ransferring available Light weight compact configuration for easy carrying Data printing or transferring to the computer 1 2 Specifications Items Specifications Air flow Test range 40 4300 m 3 h Accuracy 3 of readings 8m 3 h 85m 3 h Resolution 1 m 3 h Air velocity Test range 0 15 40 m s Pitot Tube 0 15 15 m s Velocity Grid Accuracy 3 of readings 0 05m s 0 25m s Resolution 0 01m s Differential pres...

Страница 9: ...ure 1 Outlook 2 Internal view 3 Grid Testing point Fabric Hood Micromanometer Pivot Window Frame Poles Instrument Base Poles Grid Back pressure plate Test handle Base buttom Micromanometer Back pressure ON OFF Potable handle Adjust handle ...

Страница 10: ...lineup with the top lip of the base as shown If you don t line it up correctly the hood might slide off the base when you insert the poles 2 Use the 4 positioning holes on the base stitching as a guide to ensure the hood fully wraps around the base 2 Frame Poles Front view Top view Bottom view Back Cover lock Mini USB Port Power on off Air port Air port Keypad DisplayLCD Battery Compartment Cover ...

Страница 11: ...ortable handle when packing in the carring case to avoid case damaged 4 Micromanometer basement Note 1 Correct install the micromanometer to the base with the same S N No 2 Do not cross insert the Air pines just refer to the figure as above 3 2 Velocity Grid Air velocity testing can be achieved when micromanometer work together with the velocity Grid ...

Страница 12: ...ter work together with Pitot tube 3 4 Micromanometer Differential pressure testing can be achieved by using the micromanometer Device under test 3 5 Shoulder Strap When testing by Velocity Grid Pitot tube Micromanometer as the tool for convenient view and easy operation the shoulder strap can be fit for using Static pressure port Testing point Handle Grid ...

Страница 13: ... batteries in and make certain the batteries are correctly oriented New alkaline battery or rechargeable Ni MH battery can be used Never mix battery types or battery leakage or damage to the instrument may occur Replace the compartment cover refer to figure The compartment cover will be automatically locked When power is supplied with batteries the current charge value will be displayed on the upp...

Страница 14: ...2 seconds will enter into the main MENU OK In the General Setting pressing for 2 seconds will save the date or time While in the Record Processing pressing for 2 seconds will delete saved data START Begin a measurement in main menu STOP Stop a measurement in main menu BACK Cancel or end an operation or return to the previous screen EXIT Return to the main menu ESC Stop printing when printing DOWN ...

Страница 15: ...Interface Testing information parameter settings and the testing conditions will be displayed on the main interface A Date and time real time display the current date and time B Testing data display airflow air velocity differential pressure and etc C Tools icon working tools display D Testing condition temperature humidity or absolute pressure displaying according to the Back pressure On Back pre...

Страница 16: ...y user More larger of the value of AvgTime will get more stable of the testing data 5 If auto save function On testing data will be recorded automatically according to the setting AutoSave and value of Cycle will be added automatically 6 Press STOP testing stop And STOP will be as the Left button for the Micromanometer or instrument base 7 Same testing method with the capture hood when using the V...

Страница 17: ...WN OK EXIT SET 2014 03 20 13 10 20 5 30 7 RH m3 h Atmos kPa Ready ID Cycle Mode 01 001 Avg START SAVE MENU 99 0 1 2 optional settings 1 Basic setting 2 Test setting 3 Working tool displaying below 2s 4 switch on Back Pressure Plate 5 testing ready 6 testing 2 Testing 1 Ready for testing 7 test output 8 shut off the plate 9 testing ready 2014 03 20 13 10 20 5 30 7 RH m3 h Atmos kPa Ready ID Cycle M...

Страница 18: ...te UP DOWN OK BACK SET 2014 04 01 13 10 Off High General Setting 1 2 3 4 5 Date Time Auto Shut Backlight Communicate UP DOWN OK BACK SET 2014 04 01 13 10 Off High 01 General Setting 1 2 3 4 5 Date Time Auto Shut Backlight Communicate UP DOWN OK BACK SET 2014 03 20 13 10 Off High 2s 2s 1 Press or for data modify 2 Press for hour minute selection 3 Press for 2s for save settings 4 Press for quit 1 按...

Страница 19: ...h communication for improving battery life 1 Press or for options change 2 Press for save settings 3 Press for quit 1 Press or for options change 2 Press for save settings 3 Press for quit 3 按 键 退出设置 General Setting 1 2 3 4 5 Data Time Auto Shut Backlight Communicate UP DOWN OK BACK SET 2014 03 20 13 10 Off 20 min 10 min 30 min 60 min General Setting 1 2 3 4 5 Data Time Auto Shut Backlight Communi...

Страница 20: ...Cycles UP DOWN OK BACK SET Communicate 1 2 3 Computer Printer Bluetooth UP DOWN OK BACK SET Off Communicate 1 2 3 Computer Printer Bluetooth UP DOWN OK BACK SET Off Computer Comfirm the Connection UP DOWN OK BACK SET Waiting Communicate 1 2 3 Computer Printer Bluetooth UP DOWN OK BACK SET On General Setting 1 2 3 4 5 Date Time Auto Shut Backlight Communicate UP DOWN OK BACK SET 2014 03 20 13 10 Of...

Страница 21: ... Hood Velocity Grid Pitot UP DOWN OK BACK SET 4 Pressure P Select Tool 1 2 3 Capture Hood Velocity Grid Pitot UP DOWN OK BACK SET 4 Pressure P 1 Test IDis up to 20 2 Set different Test ID for the different tool 3 100 data records can be stored in a single ID Test ID 1 2 Select ID Add New ID 01 UP DOWN OK BACK SET No Total ID 20 Test Setting 1 2 3 4 5 Select Tool Test ID Test Mode Unit Data Record ...

Страница 22: ...est Mode Unit Data Record UP DOWN OK BACK SET RunAvg 6 7 Std Act K factor 1 000 Act Test Setting 1 2 3 4 5 Select Tool Test ID Test Mode Unit Data Record UP DOWN OK BACK SET RunAvg 6 7 Std Act K factor 1 000 Std Test Setting 1 2 3 4 5 Select Tool Test ID Test Mode Unit Data Record UP DOWN OK BACK SET RunAvg 6 7 Std Act K factor Std Act Test ID UP DOWN OK BACK SET Total 20 01 Max Min Avg m3 h m3 h ...

Страница 23: ...y Model 6715 2 Under Pitot tube testing K factor is the factory standard coefficient and each pitot with its own independent factor 3 No K factor setting under the micromanometer testing 1 Press or for options change 2 Press for save settings 3 Press for quit Test Setting 1 2 3 4 5 Select Tool Test ID Test Mode Unit Data Record UP DOWN OK BACK SET RunAvg 6 7 Std Act K factor 1 000 Act 8 Temperatur...

Страница 24: ...ct cable Display DATA Error Back Pressure mode using incorrect Refer to Chapter 5 4 in Manuel Press START but no response Low battery charge Replace new batteries Back Pressure Plate opened On Shut off the plate Testing record can not be saved Over limit Cycle quantity in current ID Add new ID or delete some data New ID can not be added ID over limit Delete some ID Current ID is without Cycle Usin...

Страница 25: ...NG ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE APPLIES TO THE PRODUCT AFTER THE APPLICABLE PERIOD OF THE EXPRESS LIMITED WARRANTY STATED ABOVE AND NO OTHER EXPRESS WARRANTY OR GUARANTY EXCEPT AS MENTIONED ABOVE GIVEN BY ANY PERSON OR ENTITY WITH RESPECT TO THE PRODUCT SHALL BIND KANOMAX GROUP COMPANIES KANOMAX GROUP COMPANIES SHALL NOT BE LIABLE FOR LOSS OF STORAGE ...

Страница 26: ...a minimum period of five 5 years after termination of production This storage period of replacement parts is considered as the period during which we can provide repair service For furtherinformation please contact your local distributor or contacts on the last page PRODUCT NAME Model No Serial No Description of the problem Data of Purchase Day Month and Year When making an inquiry please provide ...

Страница 27: ...x usa com E Mail info kanomax usa com Japan Asia KANOMAX JAPAN INC 2 1 Shimizu Suita City Osaka Japan 565 0805 TEL 81 6 6877 0183 FAX 81 6 6879 2080 URL www kanomax co jp E Mail sales kanomax co jp China Shenyang Kano Scientific Instrument Co Ltd 2610 51 Wulihe Street Heping District Shenyang City PRC TEL 86 24 23845309 FAX 86 24 23898417 URL http www kanomax com cn E Mail sales kanomax com cn ...