3. Measurement
3-1 Real-time Sampling Mode
3-1-1 Measurement Preparation
1.
Place filter to Suction Nozzle.
2.
Attach Suction Nozzle to the device before measurement.
3.
Use ‘AA’ alkaline batteries or AC Adaptor.
Use AC adaptor for longer measurement.
4.
Turn on the device at clean ambient.
It takes about two minutes for the device initializing and start measuring.
At the point when initializing is done, the device sets the point as ‘Zero’ point.
To get the zero point stable, initializing at clean ambient is recommended.
Air Purifying Unit is used to set ‘Zero’ point forcibly. (Go to 3-5)
3-1-2 Measurement
1.
Turn on the device and wait for about two minutes until ‘Ready’ stops blinking and initializing is done.
2.
Start measuring.
3.
Use OMX-SRM Data Fling Software if necessary.
3-1-3 Finish Measurement
Turn off the devise when Odor Strength Level shows close to ‘0’.