NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
53
Version 1.0.1
o
Sheath
is the
Flow On/Off
status of the sheath flow applied to
the ion mobility classifier.
o
Setpoint
is the setpoint flowrate for the sheath flow applied to
the ion mobility classifier, it is automatically set according to
the selected scan mode from Kanolysis software.
▪
System Flows
allows you to configure flow settings of the Annular
Flow Ion Mobility Sizer (AFIMC + Fast CPC) system.
o
IMC Inlet
is an automatically calculated value based on the
Inlet Flow
and
Charge Conditioner Purge Flow
settings.
o
CPC Inlet
is the aerosol flow rate at the
Aerosol Inlet
of the
Fast CPC.
o
CPC Meas
is the inspected aerosol flow rate of the Fast CPC,
confirm it is set at the default value of 300 ccm for LNS
operations.
Changing the Date and/or Time
To change the date and time of AFIMC, follow these instructions:
1.
Press
F2
.
2.
Touch the
Configuration
button.
3.
On the
Additional Settings
screen, touch the
Set Date / Time
button.
Configuration
Set Date / Time