![Kanomax 6332 Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/kanomax/6332/6332_operation-manual_652714022.webp)
15
Specification
Specification
Specification
Specification of Main Unit
of Main Unit
of Main Unit
of Main Unit
Product Name
Airflow Transducer
(with / without Display)
Model
6332 / 6332D
Measuring Object
Clean Airflow
Measuring Range
Maximum range 0.1 ~ 50 m/s (Depending on the measurement range of the
probe*)
Measuring Accuracy ± (3% of r 0.1) m/s
Temperature
Compensation Range
0.1-4.99m/s
5.00-9.99m/s
10.0-24.9m/s
25.0-50.0m/s
5
~
40
℃
±
0.25m/s
±
0.50m/s
±
1.25m/s
±
2.50m/s
40
~
80
℃
±
0.35m/s
±
0.70m/s
±
1.75m/s
±
3.50m/s
Output
To select either current or voltage output (setting made with the DIP switch).
Current output: DC 4~20mA (Max. load resistance: 250
Ω
)
Voltage output: DC 0~5V
Setting of
Output Range
To be selected from the followings (setting made with the rotary switch).
0~2, 0~5, 0~10, 0~25, or 0~50 m/s
Display (Optional)
Number of digits and range of value to be displayed change in accordance
with the setting of the output range.
When setting of output range: 0~2, 0~5, 0~10 m/s
→
Display
Resolution: 0.01m/s
When setting of output range: 0~25, 0~50 m/s
→
Display Resolution:
0.1m/s
Power Supply
DC 12~24V
Power Consumption
(Reference Value)
Approx. 2.0W (Under the following conditions - power supply: 12V, air
velocity: approx. 10m/s, using a unit with a display with probe model
0965-03.)
Power consumption rate is subject to change according to conditions such as
air velocity, probe type and use of display.
Connection Wire
Size
0.5 ~ 1.5 mm
2
Operating
Temperature Range
5 ~ 40 degrees C
Storage
Temperature Range
-10 ~ 50 degrees C
Dimensions
Approx. 128(W)×78(D)×30(H)
Weight
Approx. 320g
Standard Accessory
Operation Manual : 1
Main Unit Case: 1
Optional Accessory
Probe Cable (10m, 20m, 30m)
Display Unit (Supplied with a case with a display window and 2 installation
screws.)
Dedicated AC Adapter (DC12V)
* Refer to “Specification of Probes” on page 16.